Replace The Probe Module; Check/Replace Particle Filter - TESTO 300 LL NEXT LEVEL Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 300 LL NEXT LEVEL:
Table of Contents

Advertisement

2
Blow compressed air through flue gas
ducts of the probe module and probe
handle (see illustration). Do not use a
brush!
3
Fit probe module onto the probe handle and click into place.

10.9 Replace the probe module

Disconnect the flue gas probe from the measuring instrument.
1
Press the key on the top of the probe
handle and remove the probe module.
2
Fit new probe module and click into place.

10.10 Check/replace particle filter

Check particle filter
Particle filters of the modular flue gas probe must be checked regularly for
contamination: check visually by looking through the window of the filter
chamber.
If there is visible contamination or inadequate pump flow, replace the filter.
Replace particle filter
The filter chamber may contain condensate. This is not a malfunction
and will not cause incorrect readings.
10 Maintenance
95