7 Using the product
7.2.2
Freezing a reading, displaying the
maximum/minimum value
The current reading can be frozen. The maximum and minimum values since
the last time the instrument was switched on in the standard view or during a
multi-point or continuous measurement can be displayed.
1
Press
several times until the desired value is displayed.
The following are displayed in turn:
Hold: frozen measured value
o
Max: Maximum value
o
Min: Minimum value
o
Current measuring value
o
7.2.3

Resetting maximum/minimum values

The maximum/minimum values of all channels can be reset to the current
reading.
1
Press
several times until Max or Min is shown.
2
Hold down
(approx. 2 s).
All maximum and minimum values are reset to the current reading.
7.2.4

Carrying out multi-point mean calculation

1
Press MODE/END.
flashes.
The number of readings recorded is displayed in the upper line, while
the current reading is displayed in the lower line.
2
To record readings (in the desired quantity):
Press
MENU/ENTER
3
To end measurement and calculate the mean value:
Press MODE/END.
and
flash.
16
www.
(several times).
Testo-Direct
1.888.610.7664
.com