Electrical connection
64
1.
1. Loosen the 4 fixing screws on the housing cover.
2. Open the housing cover.
3. Fold open the terminal cover.
4. Set the desired device name using the corresponding DIP switches on the
I/O electronics module.
5. Reverse the removal procedure to reassemble the transmitter.
6. Reconnect the device to the power supply.
 The configured device address is used once the device is restarted.
Setting the device name: Proline 500
Risk of electric shock when opening the transmitter housing.
Before opening the transmitter housing:
Disconnect the device from the power supply.
The default IP address may not be activated →  66.
1. Depending on the housing version, loosen the securing clamp or fixing screw of the
housing cover.
2. Depending on the housing version, unscrew or open the housing cover and
disconnect the local display from the main electronics module where necessary .
3. Set the desired device name using the corresponding DIP switches on the
I/O electronics module.
4. Reverse the removal procedure to reassemble the transmitter.
Proline Promass E 500 PROFINET
E S
-
C
+
-
2.
1 2 3 4 5 6 7 8
Name of Station
1
2
3
4
5
6
7
8
3.
4.
A0034497
128
64
32
16
8
4
2
1
A0034498
Endress+Hauser