Diagnostics and troubleshooting
186
12.6

Adapting the diagnostic information

12.6.1

Adapting the diagnostic behavior

Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for specific diagnostic information in the
Diagnostic behavior submenu.
Expert → System → Diagnostic handling → Diagnostic behavior
Available diagnostic behaviors
The following diagnostic behaviors can be assigned:
Diagnostic behavior
Alarm
Warning
Logbook entry only
Off
Displaying the measured value status
If modules with input data (e.g. Analog Input module, Discrete Input module, Totalizer
module, Heartbeat module) are configured for cyclic data transmission, the measured
value status is coded as per PROFIBUS PA Profile 3.02 Specification and transmitted along
with the measured value to the PROFINET Controller via the status byte. The status byte is
split into three segments: Quality, Quality Substatus and Limits.
Quality
 41
Structure of the status byte
The content of the status byte depends on the configured failsafe mode in the particular
function block. Depending on which failsafe mode has been configured, status information
in accordance with PROFIBUS PA Profile Specification 3.02 is transmitted to the the
PROFINET controller via the status byte. The two bits for the limits always have the value
0.
Description
The device stops measurement. The totalizers assume the defined alarm
condition. A diagnostic message is generated.
The device continues to measure. The measured value output via PROFINET
and the totalizers are not affected. A diagnostic message is generated.
The device continues to measure. The diagnostic message is displayed only in
the Event logbook submenu (Event list submenu) and not in alternation with
the operational display.
The diagnostic event is ignored, and no diagnostic message is generated or
entered.
Measuring value
Quality Substatus
Proline Promass E 500 PROFINET
Coding
(hex)
Byte 5
Limits
Endress+Hauser
A0032228-EN