Installation
24
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
tube, avoid the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical
pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the
nominal diameter prevent the sensor running empty while measurement is in progress.
7
Installation in a down pipe (e.g. for batching applications)
1
Supply tank
2
Sensor
3
Orifice plate, pipe restriction
4
Valve
5
Batching tank
DN
[mm]
8
15
25
40
50
80
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction (direction of medium flow through the piping).
Proline Promass E 500 PROFINET
1
2
3
4
5
Ø orifice plate, pipe restriction
[in]
[mm]
³⁄₈
6
½
10
1
14
1½
22
2
28
3
50
A0028773
[in]
0.24
0.40
0.55
0.87
1.10
1.97
Endress+Hauser