Endress+Hauser HART Proline Promass E 200 Operating Instructions Manual

Endress+Hauser HART Proline Promass E 200 Operating Instructions Manual

Coriolis mass flow measuring system
Hide thumbs Also See for HART Proline Promass E 200:
Table of Contents

Advertisement

Quick Links

Operating Instructions
Proline Promass E 200
HART
Coriolis mass flow measuring system
BA01027D/06/EN/13.11
71135098
Valid as of version
01.01.zz (Device software)

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser HART Proline Promass E 200

  • Page 1 Operating Instructions Proline Promass E 200 HART Coriolis mass flow measuring system BA01027D/06/EN/13.11 71135098 Valid as of version 01.01.zz (Device software)
  • Page 2 • The manufacturer reserves the right to modify technical data in line with technological developments without prior notice. Your Endress+Hauser sales center will supply you with current information and updates to this manual.
  • Page 3: Table Of Contents

    10.1 Function check ......50 10.2 Switching on the measuring device ... 50 10.3 Setting the operating language ....50 Endress+Hauser...
  • Page 4 10.4.5 Configuring the low flow cut off ..63 14.3 Endress+Hauser services ....92 10.4.6 Configuring the partial filled pipe detection .
  • Page 5: Document Information

    A connection that has to be connected to the plant grounding system: This may be a potential equalization A0011201 line or a star grounding system depending on national or company codes of practice. 1.2.3 Tool symbols Symbol Meaning Torx screwdriver A0013442 Flat blade screwdriver A0011220 Endress+Hauser...
  • Page 6: Symbols For Certain Types Of Information

    Meaning 1, 2, 3,... Item numbers Series of steps … A, B, C, ... Views A-A, B-B, C-C, ... Sections Flow direction A0013441 Hazardous area Indicates a hazardous area. A0011187 Safe area (non-hazardous area) Indicates a non-hazardous area. A0011188 Endress+Hauser...
  • Page 7: Documentation

    The document types listed are available: • On the CD supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download 1.3.2 Supplementary device-dependent documentation If operating the device in SIL applications, hazardous areas or in accordance with the Pressure Equipment Directive, comply strictly with the instructions in the supplementary documentation.
  • Page 8: Basic Safety Instructions

    Observe the specified maximum process pressure. ► Verification for borderline cases: For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying ► the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability.
  • Page 9: Workplace Safety

    It fulfills general safety requirements and legal requirements. It also conforms to the EC directives listed in the device-specific EC declaration of conformity. Endress+Hauser confirms this fact by applying the CE mark. Endress+Hauser...
  • Page 10: Product Description

    Registered trademark of Ladish & Co., Inc., Kenosha, USA Ò Ò Ò Applicator , FieldCare , Field Xpert , HistoROM Registered or registration-pending trademarks of the Endress+Hauser Group Patents US patents EP patents 5,610,342 0 701 213 6,513,393 1 055 102...
  • Page 11 Proline Promass E 200 HART Product description US patents EP patents 6,412,355 1 054 240 7,555,397 Endress+Hauser...
  • Page 12: Incoming Acceptance And Product

    DELIVERY NOTE A0015502 A0013699 Do the nameplate data match the ordering information on the delivery note? A0015502 A0013697 Is the CD-ROM with the Technical Documentation and documents present? If one of the conditions does not comply, contact your Endress+Hauser distributor. Endress+Hauser...
  • Page 13: Product Identification

    Additional information about the transmitter version (certificates, approvals), e.g. SIL, PROFIBUS Permitted temperature range for cable Manufacturing date: year-month Degree of protection: e.g. IP, NEMA Approval information for explosion protection Document number of safety-related supplementary documentation, e.g. Safety Instructions 2-D matrix code Endress+Hauser...
  • Page 14 (e.g. LA). If other optional specifications are also ordered, these are indicated collectively using the # placeholder symbol (e.g. #LA#). • If the ordered optional specifications do not include any safety and approval-related specifications, they are indicated by the + placeholder symbol (e.g. 8E2B50-ABCDE+). Endress+Hauser...
  • Page 15: Storage And Transport

    • For measuring device > DN 40 (1½ in): lift the measuring device using the webbing slings at the process connections; do not lift at the transmitter housing. • Do not remove protective covers or protective caps installed on process connections. They prevent mechanical damage to the sealing surfaces and fouling in the measuring tube. Endress+Hauser...
  • Page 16: Packaging Disposal

    • Seaworthy packaging (optional): Wood crate, treated in accordance with ISPM 15 standard, which is confirmed by the affixed IPPC logo. • Carrying and mounting hardware: – Disposable plastic pallet – Plastic straps – Plastic adhesive strips • Dunnage: Paper cushion Endress+Hauser...
  • Page 17: Mounting

    Notwithstanding the above, the following installation proposal permits installation in an open down pipe. Pipe restrictions or the use of an orifice plate with a smaller cross-section than the nominal diameter prevent the sensor running empty while measurement is in progress. Endress+Hauser...
  • Page 18 Exception: A0015589 Horizontal orientation, transmitter head down Exception: A0015590 Horizontal orientation, transmitter head at side A0015592 Applications with high process temperatures can increase the ambient temperature. This orientation is recommended to observe the maximum ambient temperature for the transmitter. Endress+Hauser...
  • Page 19: Process

    Cavitation is caused if the pressure drops below the vapor pressure: • In liquids that have a low boiling point (e.g. hydrocarbons, solvents, liquefied gases) • In suction lines Ensure the system pressure is sufficiently high to prevent cavitation and outgassing. ► Endress+Hauser...
  • Page 20 For this reason, the sensor must be magnetically shielded: the housing can be shielded with tin plates or electric sheets without a privileged direction (e.g. V330-35A). The sheet must have the following properties: • Relative magnetic permeability mr ³ 300 • Plate thickness d ³ 0.35 mm (d ³ )0.014 in Endress+Hauser...
  • Page 21: Special Mounting Instructions

    Preparing the measuring device 1. Remove all remaining transport packaging. 2. Remove any protective covers or protective caps present from the sensor. 3. If present, remove transport protection of the rupture disk. 4. Remove stick-on label on the electronics compartment cover. Endress+Hauser...
  • Page 22: Mounting The Measuring Device

    350° 8 mm 8 mm A0013713 1. Unscrew the securing screw using an open-ended wrench. 2. Rotate the housing in the desired direction. 3. Firmly tighten the securing screw. 6.2.5 Turning the display module 3 mm A0013905 Endress+Hauser...
  • Page 23: Post-Mounting Check

    Does the arrow on the sensor nameplate match the direction of flow of the fluid through the piping (® ä 18)? Are the measuring point identification and labeling correct (visual inspection)? Is the device adequately protected from precipitation and direct sunlight? Are the securing screw and securing clamp tightened securely? Endress+Hauser...
  • Page 24: Electrical Connection

    Have electrical connection work carried out by correspondingly trained specialists only. ► Observe applicable federal/national installation codes and regulations. ► Comply with local workplace safety regulations. ► For use in potentially explosive atmospheres, observe the information in the device-specific Ex ► documentation. Endress+Hauser...
  • Page 25: Connecting The Signal Cables

    7. NOTICE! Housing degree of protection voided due to insufficient sealing of the housing. Screw in the thread without using lubricant. The cover threads are coated with a dry lubricant. Screw the connection compartment cover closed and tighten the securing clamp. Endress+Hauser...
  • Page 26: Connecting To The Supply Unit

    - 18 V) : 0.0036 A = 18 to 18.9 V: R • For U £ (U - 13.5 V) : 0.022 A = 18.9 to 24.5 V: R • For U £ 500 W = 24.5 to 30 V: R Endress+Hauser...
  • Page 27: Ensuring The Degree Of Protection Of The Measuring Device

    Are cables or the device damaged (visual inspection)? Do the cables comply with the requirements (® ä 24)? Do the cables have adequate strain relief? Are all cable glands installed, firmly tightened and correctly sealed? Cables looped as "water traps" (® ä 27)? Endress+Hauser...
  • Page 28 Is the terminal assignment correct (® ä 25)? If supply voltage is present, is the device ready for operation and do values appear on the display module? Are all housing covers installed and firmly tightened? Is the securing clamp tightened correctly? Endress+Hauser...
  • Page 29: Operation Options

    Proline Promass E 200 HART Operation options Operation options Overview of operation options A0015607 Local operation via display module Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX100 Field Communicator 375, 475 PLC (programmable logic controller) Endress+Hauser...
  • Page 30: Structure And Function Of The Operating Menu

    Parameter 1 Parameter n Submenu 1 Submenu n Parameter 1 Diagnostics Parameter n Submenu 1 Submenu n Operating menu for experts Direct access Expert Locking status Access status display System Sensor Output 1 Output n Communication Application Diagnostics A0013431-EN Endress+Hauser...
  • Page 31: Operating Philosophy

    Contains all parameters for configuring the digital communication interface. • "Application" submenu Contains all parameters for configuring the functions that go beyond the actual measurement (e.g. totalizer). • "Diagnostics" submenu Contains all parameters for error detection and analyzing process and device errors. Endress+Hauser...
  • Page 32: Access To The Operating Menu Via The Local Display

    The measuring device is hardware locked (® ä 74). A0013963 Communication symbols The measuring device communicates via HART protocol. A0013965 The symbol for event level and status signal of the diagnostic event that currently has highest priority for the measuring device are displayed. Endress+Hauser...
  • Page 33 Reduces the contrast (brighter setting). A0013953 Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). A0013954 Minus/Plus/Enter key combination (press and hold down the keys simultaneously) Enables or disables the keypad lock. A0013955 Endress+Hauser...
  • Page 34: Navigation View And Operating Elements

    If a diagnostic event is present, the symbol for event level and status signal Tip for the event level symbol and status signal (® ä 32) About the function and input of the direct access code refer to the document "Description of Device Parameters" Endress+Hauser...
  • Page 35 • Pressing the key briefly – Opens the selected menu, submenu or parameter. – Starts the wizard. A0013952 • Pressing the key for 2 s for parameter: If present, opens the help text for the function of the parameter. Endress+Hauser...
  • Page 36: Editing View And Operating Elements

    The following input symbols are available in the input mask of the numeric and text editor: Numeric editor symbols Selection of numbers from 0 to 9 Inserts decimal separator or minus .– sign at the input position. … A0013984 A0013998 Endress+Hauser...
  • Page 37 A0013971 Operating functions in the numeric editor Minus/Enter key combination (press and hold down the keys simultaneously) Reduces the contrast (brighter setting). A0013953 Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). A0013954 Endress+Hauser...
  • Page 38: Opening The Context Menu

    Ã The context menu is closed and the measured value display appears. Calling up the menu via the context menu 1. Open the context menu. 2. Press O to navigate to the desired menu. 3. Press F to confirm the selection. Ã The selected menu opens. Endress+Hauser...
  • Page 39: Navigating And Selecting From List

    For some parameters, help texts exist, which the user can call up from the navigation view. These briefly describe the function of the parameter and thus support fast and reliable commissioning. Calling up and closing the help text The user is in the navigation view and the selection bar is on a parameter. Endress+Hauser...
  • Page 40 Ã The help text for the selected parameter opens. Ent. access code Enter access code to disable write protec. A0014002-EN å 8 Example: Help text for parameter "Enter access code" 2. Press S + O simultaneously. Ã The help text is closed. Endress+Hauser...
  • Page 41: Changing The Parameters

    100.000 kg/s 4 mA value 0.000 kg/s 0.000 kg/s kg/s 100.000 kg/s 100.000 kg/s kg/s kg/s kg/s kg/s kg/s Curr. output 1 Failure mode Max. Curr. output 1 20 mA value 20.000 kg/s 0.000 kg/s 4 mA value A0014020-EN Endress+Hauser...
  • Page 42: User Roles And Related Access Authorization

    Only the measured values on the measured value display can be read off. The keypad lock is enabled and disabled in the same way: The user is in the measured value display. Endress+Hauser...
  • Page 43 Ã After enabling the keypad lock: XXXXXXXXXX 20.50 Keylock on kg/h A0014000-EN After disabling the keypad lock: XXXXXXXXXX 20.50 Keylock off kg/h A0014001-EN If the user attempts to access the operating menu while the keylock is enabled, the message "Keylock on" also appears. Endress+Hauser...
  • Page 44: Access To The Operating Menu Via The Operating Tool

    FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and help you manage them. By using the status information, it is also a simple but effective way of checking their status and condition. Access takes place via HART protocol or CDI interface.
  • Page 45: Ams Device Manager

    HART protocol. Source for device description files See data (® ä 48) 8.4.5 Field Communicator 475 Function scope Compact, flexible and robust industrial handheld terminal from Emerson Process Management for remote configuration and measured value display via HART protocol. Endress+Hauser...
  • Page 46: Connecting Operating Tools

    Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475 Field Communicator 375, 475 Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM) Commubox FXA191 (RS232) or FXA195 (USB) Field Xpert SFX100 VIATOR Bluetooth modem with connecting cable Transmitter Endress+Hauser...
  • Page 47 Proline Promass E 200 HART Operation options Via service interface (CDI) A0014019 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface) Commubox FXA291 Computer with "FieldCare" operating tool Endress+Hauser...
  • Page 48: System Integration

    Sources for obtaining device descriptions Field Xpert SFX100 Use update function of handheld terminal FieldCare • ® Download Area www.endress.com • CD–ROM (contact Endress+Hauser) • DVD (contact Endress+Hauser) AMS Device Manager ® Download Area www.endress.com (Emerson Process Management) SIMATIC PDM ®...
  • Page 49: Other Settings

    In the HART configuration submenu, you can configure other settings for the HART protocol (e.g. Burst mode). Navigation path: Expert ® Communication ® HART config. For an overview and description of the parameters of this menu, refer to the document "Description of Device Parameters" Endress+Hauser...
  • Page 50: Commissioning

    Factory setting: English or ordered local language X X X X X X X 20.50 Main menu 0104-1 Language English Display/operat. Setup Language 0104-1 English à Deutsch Español Français Language 0104-1 à English Deutsch Español Français Hauptmenü 0104-1 Sprache Deutsch Anzeige/Betrieb Setup A0013996 Endress+Hauser...
  • Page 51: Configuring The Measuring Device

    Curr. output 1 ® (® ä 54) Curr. output 2 ® (® ä 54) ® Display (® ä 58) Output conduct ® (® ä 60) Low flow cut off ® (® ä 63) Partial filled pipe ® (® ä 64) detection Endress+Hauser...
  • Page 52: Selecting And Setting The Medium

    "Select medium" wizard in the "Setup" menu Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Select medium Select the medium type. • Liquid Liquid • Gas Select gas type Select the gas type for the Gas type choose list measurement application. Endress+Hauser...
  • Page 53 Pressure compensation Enable the automatic • Off pressure correction. • Fixed Pressure value Enter a value for the process 0 to 99 999 [bar, psi] Country-dependent: pressure to be used for • 1.01325 bar pressure correction. • 14.7 psi Endress+Hauser...
  • Page 54: Configuring The Current Outputs

    Fixed current NAMUR 4 mA value 20 mA value Failure mode Min. Defined value Failure current End of wizard A0013796-EN å 11 "Current output 1-2" in the "Setup" menu (Part 1) Off* = Option only for current output 2 Endress+Hauser...
  • Page 55 Current span 4...20 mA 4...20 mA US 4...20 mA Fixed current NAMUR 4 mA value 20 mA value Failure mode Min. Defined value Failure current End of wizard A0013844-EN å 12 "Current output 1-2" in the "Setup" menu (Part 2) Endress+Hauser...
  • Page 56 • lb/scf Result The selected unit applies for: – Current outputs – Low value partial filled pipe detection – High value partial filled pipe detection – Simulation process variable – Fixed reference density – Density adjustment (in the Expertmenu) Endress+Hauser...
  • Page 57 Max. output adopts in an alarm • Max. condition. • Last valid value • Actual value • Defined value Failure current Select the current value the 3.59 to 22.5 mA 22.5 mA current output adopts in an alarm condition. Endress+Hauser...
  • Page 58: Configuring The Local Display

    Value 1 display 0% bargraph value 1 100% bargraph value 1 Value 2 display Value 3 display 0% bargraph value 3 100% bargraph value 3 Value 4 display End of wizard A0013797-EN å 13 "Display" wizard in the "Setup" menu Endress+Hauser...
  • Page 59 • Current output 1 • Current output 2 0% bargraph value 3 Enter the 0% value to be Max. 15-digit floating-point Country-dependent: shown on the bargraph number with sign • 0 kg/h display for the measured • 0 lb/min value 3. Endress+Hauser...
  • Page 60: Configuring The Output Conditioning

    • Current output 1 • Current output 2 10.4.4 Configuring the output conditioning The Output conditioning wizard guides you systematically through all parameters that have to be set for configuring the output conditioning. Navigation path "Setup" menu ® Output conditioning Endress+Hauser...
  • Page 61 On value low flow cutoff Off value low flow cutoff Pressure shock suppression End of wizard A0013908-EN å 14 "Output conditioning" wizard in the "Setup" menu The "Reverse flow compensation" option was selected in the "Measuring mode output 1" parameter Endress+Hauser...
  • Page 62 0 to 100 % 50 % Off value low flow cut off Enter the off-value for the low flow cut off. Pressure shock suppression Enter the time interval for 0 to 100 s signal suppression (= active pressure shock suppression). Endress+Hauser...
  • Page 63: Configuring The Low Flow Cut Off

    0 to 100 % 50 % Off value low flow cut off Enter the off-value for the low flow cut off. Pressure shock suppression Enter the time interval for 0 to 100 s signal suppression (= active pressure shock suppression). Endress+Hauser...
  • Page 64: Configuring The Partial Filled Pipe Detection

    • 6 kg/l • 374.6 lb/cf Response time part. filled pipe Enter the time interval until the 0 to 100 s detect. diagnostic message S862 Partly filled pipe detection is displayed for an empty or partially filled pipe. Endress+Hauser...
  • Page 65: Advanced Settings

    Define access code (® ä 73) Device tag (® ä 66) System units ® (® ä 66) Sensor adjustment ® (® ä 68) Totalizer 1 ® (® ä 69) Totalizer 2 ® (® ä 69) ® Totalizer 3 (® ä 69) Endress+Hauser...
  • Page 66: Defining The Tag Name

    "Setup" menu ® Advanced setup ® System units Structure of the submenu ® System units Mass flow unit Mass unit Volume flow unit Volume unit Corrected volume flow unit Corrected volume unit Density unit Reference density unit Temperature unit Length unit Endress+Hauser...
  • Page 67 Country-dependent • kg/l Result • lb/cf The selected unit applies for: – Current outputs – Low value partial filled pipe detection – High value partial filled pipe detection – Simulation process variable – Density adjustment (in the Expert menu) Endress+Hauser...
  • Page 68: Carrying Out A Sensor Adjustment

    Change the sign of the • Flow in arrow direction Flow in arrow direction direction of flow of the fluid. • Flow against arrow direction Zero point adjustment Start the zero point • Cancel Cancel control adjustment. • Start Endress+Hauser...
  • Page 69: Configuring The Totalizer

    One of the following options is Define how the • Stop Stop selected in the Assign process totalizer behaves in • Actual value variable parameter: an alarm condition. • Last valid value • Mass flow • Volume flow • Corrected volume flow Endress+Hauser...
  • Page 70: Carry Out Additional Display Configurations

    (® ä 59) 0% bargraph value 3 (® ä 59) 100% bargraph value 3 (® ä 59) Decimal places 3 Value 4 display (® ä 59) Decimal places 4 Display interval Display damping (® ä 56) Header Header text Separator Endress+Hauser...
  • Page 71: Configuration Management

    The last backup copy of the device configuration is copied from the display module to the HistoROM of the device. The backup copy comprises the transmitter data of the device. Duplicate The transmitter configuration from another device is duplicated to the device using the display module. Endress+Hauser...
  • Page 72: Simulation

    Navigation path "Diagnostics" menu ® Simulation Structure of the submenu Simulation ® Assign simulation process variable Value process variable Simulation current output 1 Value current output 1 Simulation current output 2 Value current output 2 Simulation device alarm Endress+Hauser...
  • Page 73: Protecting Settings From Unauthorized Access

    Define access code 1. Navigating to the "Define access code" parameter: Setup ® Advanced setup ® Def. access code 2. Define a max. 4-digit numeric code as an access code. Ã The -symbol appears in front of all write-protected parameters. Endress+Hauser...
  • Page 74: Write Protection Via Lock Switch

    - other than the Contrast display parameter - to be locked. The values of the parameters are still visible, but can no longer be changed (except for Contrast display), either via the local display, CDI interface or HART protocol. XXXXXXXXXXXXX 3 mm A0013768 Endress+Hauser...
  • Page 75 6. Feed the spiral cable into the gap between the housing and main electronics module and plug the display module into the electronics compartment in the desired direction until it engages. 7. Screw the electronics compartment cover closed and tighten the securing clamp. Endress+Hauser...
  • Page 76: Operation

    Mass flow Volume flow Corrected volume flow Density Reference density Temperature "Totalizer" submenu Totalizer ® Totalizer value 1 Totalizer overflow 1 Totalizer value 2 Totalizer overflow 2 Totalizer value 3 Totalizer overflow 3 "Output value" submenu Output value ® Endress+Hauser...
  • Page 77: Adapting The Measuring Device To The Process Conditions

    Preset parameter. Reset + totalize The totalizer is reset to 0 and the totaling process is restarted. Preset + totalize The totalizer is set to the defined start value in the Preset parameter and the totaling process is restarted. Endress+Hauser...
  • Page 78: Show Data Logging

    Chart of a measured value trend • x-axis: depending on the number of channels selected displays 250 to 1000 measured values of a process variable. • y-axis: displays the approximate measured value span and constantly adapts this to the ongoing measurement. Endress+Hauser...
  • Page 79 Proline Promass E 200 HART Operation If the length of the logging interval or the assignment of the process variables to the channels is changed, the content of the data logging is deleted. Endress+Hauser...
  • Page 80: Diagnostics And Troubleshooting

    • FXA 195 HART: Document "Technical Information" TI00404F No connection via service interface Incorrect configuration of USB interface Observe the documentation for the (CDI) on computer or PC or driver not installed Commubox. correctly. FXA 291 HART: Document "Technical Information" TI00405C Endress+Hauser...
  • Page 81: Diagnostic Information On Local Display

    Symbols for event level "Alarm" The measurement is interrupted. The signal outputs take on the defined alarm condition. A diagnostic message is generated. A0013961 "Warning" The device continues to measure. The signal outputs are not affected. A diagnostic message is generated. A0013962 Endress+Hauser...
  • Page 82: Calling Up Remedial Measures

    Curr. output 1 (ID:153) S441 2d11h05m55s 1. Check process 2. Check current output settings A0013940-EN Event text Symbol for event level with diagnostic event Event ID for service Operation time of occurrence Remedial measures The user is in the diagnostic message. Endress+Hauser...
  • Page 83: Diagnostic Event In The Operating Tool

    In the Diagnostics list submenu, up to 5 currently pending diagnostic messages can be displayed. If more than 5 messages are pending, the messages with the highest priority are shown on the display. Navigation path Diagnostics ® Diagnostics list Endress+Hauser...
  • Page 84: Overview Of Diagnostic Events

    The device continues to measure. The diagnostic message is entered in the Event logbook (events list) submenu only and not displayed in alternation with the measured value display. Ignore The diagnostic event is ignored, and no diagnostic message is generated or entered. Endress+Hauser...
  • Page 85: Diagnostics For The Sensor

    I/O module Change I/O module. Alarm F276 I/O module 1. Restart device. Alarm 2. Change I/O module. F282 Data storage 1. Restart device. Alarm 2. Contact service. F283 Memory content 1. Transfer data or reset device. Alarm 2. Contact service. Endress+Hauser...
  • Page 86: Diagnostics For The Configuration

    Increase the ambient temperature around the Warning range sensor housing. S832 Ambient temperature Electronics temperature is too high! Warning range Reduce ambient temperature S833 Ambient temperature Electronics temperature is too low! Warning range Increase ambient temperature. S834 Process temperature Reduce process temperature. Warning Endress+Hauser...
  • Page 87: Event Logbook

    A maximum of 20 event messages can be displayed in chronological order. If the advanced HistoROM function is enabled in the device (order option), up to 100 entries can be displayed. The event history includes entries for: • Diagnostic events (® ä 84) • Information events (® ä 88) Endress+Hauser...
  • Page 88: Filtering The Event Logbook

    Configuration modified I1092 Data logging cleared I1110 Write protection switch changed I1111 Density adjust. error I1137 Electronics changed I1151 History reset I1154 Minimum/maximum terminal voltage reset I1155 Electronics temperature reset I1156 Trend block memory error I1157 Memory content events list Endress+Hauser...
  • Page 89: Firmware History

    "Manufacturer's information" document. The manufacturer's information is available: • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download • Specify the following details: – Product root, e.g. 8E2B –...
  • Page 90: Repair

    • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices by Endress+Hauser Service or at the factory only.
  • Page 91: Endress+Hauser Services

    Proline Promass E 200 HART Repair 13.3 Endress+Hauser services For information about service and spare parts, contact your Endress+Hauser distributor. Endress+Hauser...
  • Page 92: Maintenance

    • Observe the maximum permitted fluid temperature for the measuring device (® ä 101). 14.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser representative can provide detailed information on the services.
  • Page 93: Return

    Proline Promass E 200 HART Return Return Observe the following points for returning the device: • Contact your Endress+Hauser Sales Center to obtain information about the procedure and basic conditions. • Enclose the completed "Decontamination declaration" form with the device. The form is available:...
  • Page 94: Disposal

    Ensure that the measuring device and all cavities are free of fluid residues that are hazardous to ► health or the environment, e.g. substances that have permeated into crevices or diffused through plastic. Observe the following notes during disposal: • Observe valid federal/national regulations. • Ensure proper separation and reuse of the device components. Endress+Hauser...
  • Page 95: Technical Data

    0 to 1 650 0 to 70 000 0 to 2 570 Measuring ranges for gases The full scale values depend on the density of the gas and can be calculated with the formula below: × r max(G) max(F) Endress+Hauser...
  • Page 96: Output

    • Local display – Status signal (in accordance with NAMUR Recommendation NE 107) – Plain text display • Operating tool via HART communication or service interface (CDI) – Status signal (in accordance with NAMUR Recommendation NE 107) – Plain text display Endress+Hauser...
  • Page 97: Power Supply

    • Error messages (incl. total operated hours) are stored. Electrical connection (® ä 24) Potential equalization No special measures for potential equalization are required. Terminals Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm (20 to 14 AWG) Endress+Hauser...
  • Page 98: Performance Characteristics

    Temperature ±0.5 °C ± 0.005 × T °C (±0.9 °F ± 0.005 × (T – 32) °F) Zero point stability Zero point stability [mm] [in] [kg/h] [lb/min] 0.24 0.0088 ½ 0.78 0.0287 2.16 0.0794 1½ 5.40 0.1985 8.40 0.3087 Endress+Hauser...
  • Page 99 ±0.25 °C ± 0.0025 × T °C (±0.45 °F ± 0.0025 × (T–32) °F) Response time • The response time depends on the configuration (damping). • Response time in the event of erratic changes in the measured variable: after 500 ms 95 % of the full scale value Endress+Hauser...
  • Page 100: Installation

    17.8 Environment Ambient temperature range Measuring device –40 to +60 °C (–40 to +140 °F) Local display –20 to +60 °C (–4 to +140 °F), the readability of the display may be impaired at temperatures outside the temperature range. Endress+Hauser...
  • Page 101: Process

    Rupture disks cannot be combined with the separately available heating jacket (® ä 108). Flow limit Select the nominal diameter by optimizing between the required flow range and permissible pressure loss. For an overview of the measuring range full scale values, see the "Measuring range" section (® ä 95) Endress+Hauser...
  • Page 102 2.10× 10 ½ –3 8.30× 10 7.62× 10 1.73× 10 2.13× 10 –3 12.00× 10 1.89× 10 4.66× 10 6.11× 10 1½ –3 17.60× 10 4.42× 10 1.35× 10 1.38× 10 26.00× 10 –3 8.54× 10 4.02× 10 2.31× 10 Endress+Hauser...
  • Page 103: 17.10 Mechanical Construction

    Weight Compact version All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg] Weight [kg] [mm] Order characteristic for "Housing", Order characteristic for "Housing" option C option B Aluminum coating 1.4404/316L 10.5 15.5 24.5 Endress+Hauser...
  • Page 104 • Ex ic • Ex nA • Ex t Thread G ½" For non-Ex and Ex Stainless steel 1.4404/316L via adapter (except for CSA Ex d/XP) Thread ½" NPT For non-Ex and Ex via adapter Thread M20 × 1.5 Ex d Endress+Hauser...
  • Page 105: 17.11 Operability

    • Format for displaying measured values and status variables can be individually configured • Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F) The readability of the display may be impaired at temperatures outside the temperature range. Endress+Hauser...
  • Page 106: 17.12 Certificates And Approvals

    Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 107: 17.13 Application Packages

    The Manual for Functional Safety with all information about the SIL device is available from your Endress+Hauser distributor. Pressure Equipment Directive • With the PED/G1/III marking on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure Equipment Directive 97/23/EC.
  • Page 108: 17.14 Accessories

    Technical data Proline Promass E 200 HART 17.14 Accessories For an overview of the accessories that can be ordered, see the "Technical Information" document. Endress+Hauser...
  • Page 109: Index

    Commubox FXA291 ......47 Degree of protection ......27 Endress+Hauser...
  • Page 110 Device variables ......48 Endress+Hauser services Measured values ......48 About maintenance .
  • Page 111 Performance characteristics ..... . 98 System pressure ......19 Endress+Hauser...
  • Page 112 ........77 For locking ......32 Endress+Hauser...
  • Page 113 Vibration resistanc ......101 Vibrations ........21 Endress+Hauser...
  • Page 115 Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
  • Page 116 www.endress.com/worldwide BA01027D/06/EN/13.11 71135098 EH-COSIMA ProMoDo...

Table of Contents