Diagnostic information
No.
022 Temperature sensor defective
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Influenced measured variables
• Oscillation amplitude 1
• Oscillation amplitude 2
• Application specific output
• Application specific output
• Signal asymmetry
• Carrier mass flow
• Carrier pipe temperature
• Target corrected volume flow
• Carrier corrected volume flow
• Concentration
• Oscillation damping 1
• Oscillation damping 2
• Density
• Oil density
• Water density
• Dynamic viscosity
• Sensor electronic temperature (ISEM)
• GSV flow
Endress+Hauser
12.7

Overview of diagnostic information

• The amount of diagnostic information and the number of measured variables
affected increase if the measuring device has one or more application packages.
• All of the measured variables affected in the entire Promass instrument family are
always listed under "Measured variables affected". The measured variables available
for the device in question depend on the device version. When assigning the
measured variables to the device functions, for example to the individual outputs, all
of the measured variables available for the device version in question are available
for selection.
In the case of some items of diagnostic information, the diagnostic behavior can be
changed. Change the diagnostic information →  186
12.7.1

Diagnostic of sensor

Short text
Bad
Maintenance alarm
0x24 to 0x27
F
Alarm
• GSV flow alternative
• Kinematic viscosity
• Mass flow
• Oil mass flow
• Water mass flow
• Index inhomogeneous medium
• Index suspended bubbles
• HBSI
• NSV flow
• NSV flow alternative
• External pressure
• Exciter current 1
• Exciter current 2
• Oscillation frequency 1
• Oscillation frequency 2
• S&W volume flow
• Reference density
• Reference density alternative
Remedy instructions
1. Check or replace sensor electronic module (ISEM)
2. If available: Check connection cable between sensor and transmitter
3. Replace sensor
• Corrected volume flow
• Oil corrected volume flow
• Water corrected volume flow
• Oscillation damping fluctuation 1
• Oscillation damping fluctuation 2
• Frequency fluctuation 1
• Frequency fluctuation 2
• Target mass flow
• Carrier volume flow
• Target volume flow
• Temp. compensated dynamic viscosity
• Temp. compensated kinematic viscosity
• Temperature
• Status
• Volume flow
• Oil volume flow
• Water volume flow
• Water cut
Diagnostics and troubleshooting