Electrical connection
50
Integrating into a ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and
the connection to the service interface (CDI-RJ45).
Note the following when connecting:
• Recommended cable: CAT5e, CAT6 or CAT7, with shielded connector (e.g. brand:
YAMAICHI ; Part No Y-ConProfixPlug63 / Prod. ID: 82-006660)
• Maximum cable thickness: 6 mm
• Length of connector including bend protection: 42 mm
• Bending radius: 2.5 x cable thickness
1
1
PROFINET connection
2
Service interface (CDI-RJ45)
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option NB: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted
in the cable entry. Therefore the connection to the service interface can be established
via an M12 connector without opening the device.