Endress+Hauser Proline Promass E 500 PROFINET Operating Instructions Manual page 187

Coriolis flowmeter
Table of Contents

Advertisement

Proline Promass E 500 PROFINET
Endress+Hauser
Supported status information
Status
BAD - Maintenance alarm
BAD - Process related
BAD - Function check
UNCERTAIN - Initial value
UNCERTAIN - Maintenance demanded
UNCERTAIN - Process related
GOOD - OK
GOOD - Maintenance demanded
GOOD - Function check
Determining the measured value status and device status via the diagnostic behavior
When the diagnostic behavior is assigned, this also changes the measured value status and
device status for the diagnostic information. The measured value status and device status
depend on the choice of diagnostic behavior and on the group in which the diagnostic
information is located.
The diagnostic information is grouped as follows:
• Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
→  187
• Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
→  187
• Diagnostic information pertaining to the configuration: diagnostic number 400 to 599
→  188
• Diagnostic information pertaining to the process: diagnostic number 800 to 999
→  188
Depending on the group in which the diagnostic information is located, the following
measured value status and device status are firmly assigned to the particular diagnostic
behavior:
Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
Diagnostic behavior
(configurable)
Quality
Alarm
BAD
Warning
GOOD
Logbook entry only
GOOD
Off
Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
Diagnostic number 200 to 301, 303 to 399
Diagnostic behavior
(configurable)
Quality
Alarm
BAD
Warning
Coding (hex)
0x24
0x28
0x3C
0x4F
0x68
0x78
0x80
0xA8
0xBC
Measured value status (fixed assignment)
Quality
Coding
Substatus
(hex)
Maintenance
0x24
alarm
Maintenance
0xA8
demanded
ok
0x80
Measured value status (fixed assignment)
Quality
Coding
Substatus
(hex)
Maintenance
0x24
alarm
Diagnostics and troubleshooting
Device diagnosis
Category
(fixed assignment)
(NE107)
F
Maintenance
(Failure)
alarm
M
Maintenance
(Maintenance)
demanded
Device diagnosis
Category
(fixed assignment)
(NE107)
F
Maintenance
(Failure)
alarm
187