Endress+Hauser Proline Promass E 500 PROFINET Operating Instructions Manual page 38

Coriolis flowmeter
Table of Contents

Advertisement

Electrical connection
38
Cross-section
Cable length [max.]
2 x 2 x 0.50 mm
2
50 m (165 ft)
(AWG 20)
3 x 2 x 0.50 mm
2
100 m (330 ft)
(AWG 20)
2
4 x 2 x 0.50 mm
150 m (500 ft)
(AWG 20)
Optionally available connecting cable
Connecting cable for
Zone 1; Class I, Division 1
Standard cable
2 × 2 × 0.5 mm
stranded)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Operating temperature
When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable
can move freely: –25 to +105 °C (–13 to +221 °F)
Available cable length
Fixed: 20 m (65 ft); variable: up to maximum 50 m (165 ft)
1)
UV radiation can impair the cable outer sheath. Protect the cable from direct sunshine where possible.
Proline Promass E 500 PROFINET
Termination
2 x 2 x 0.50 mm
2
(AWG 20)
BN WT YE GN
GY
• +, – = 0.5 mm
2
• A, B = 0.5 mm
2
3 x 2 x 0.50 mm
2
(AWG 20)
BN
WT GY PK
GY
• +, – = 1.0 mm
2
• A, B = 0.5 mm
2
2
4 x 2 x 0.50 mm
(AWG 20)
BN
WT GY PK
GY
2
• +, – = 1.5 mm
2
• A, B = 0.5 mm
1)
2
(AWG 20) PVC cable
with common shield (2 pairs, pair-
A
B
YE GN
A
B
RD BU
A
B
YE GN
Endress+Hauser

Advertisement

Table of Contents
loading

Table of Contents