Diagnostics and troubleshooting
Diagnostic information
No.
438 Dataset
Measured variable status
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Influenced measured variables
• Oscillation amplitude 1
• Oscillation amplitude 2
• Application specific output
• Application specific output
• Signal asymmetry
• Carrier mass flow
• Carrier pipe temperature
• Target corrected volume flow
• Carrier corrected volume flow
• Concentration
• Measured values 1
• Measured values 2
• Measured values 3
• Oscillation damping 1
• Oscillation damping 2
• Density
• Oil density
• Water density
• Dynamic viscosity
Diagnostic information
No.
441 Current output 1 to n
Measured variable status [from the factory]
Quality
Quality substatus
Coding (hex)
Status signal
Diagnostic behavior
Influenced measured variables
–
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
218
Short text
Uncertain
Maintenance demanded
0x68 to 0x6B
M
Warning
• Sensor electronic temperature (ISEM)
• GSV flow
• GSV flow alternative
• Kinematic viscosity
• Mass flow
• Oil mass flow
• Water mass flow
• Index inhomogeneous medium
• Index suspended bubbles
• HBSI
• NSV flow
• NSV flow alternative
• External pressure
• Exciter current 1
• Exciter current 2
• Oscillation frequency 1
• Oscillation frequency 2
• S&W volume flow
• Reference density
Short text
1)
Good
Ok
0x80 to 0x83
S
Warning
Remedy instructions
1. Check data set file
2. Check device configuration
3. Up- and download new configuration
• Reference density alternative
• Corrected volume flow
• Oil corrected volume flow
• Water corrected volume flow
• Oscillation damping fluctuation 1
• Oscillation damping fluctuation 2
• Frequency fluctuation 1
• Frequency fluctuation 2
• Target mass flow
• Carrier volume flow
• Target volume flow
• Temp. compensated dynamic viscosity
• Temp. compensated kinematic viscosity
• Temperature
• Status
• Volume flow
• Oil volume flow
• Water volume flow
• Water cut
Remedy instructions
1. Check process
2. Check current output settings
Proline Promass E 500 PROFINET