16.12 Certificates And Approvals - Endress+Hauser Proline Promass E 500 PROFINET Operating Instructions Manual

Coriolis flowmeter
Table of Contents

Advertisement

Technical data
CE mark
Ex approval
Sanitary compatibility
278
Event list
Automatic
• Chronological display of up to 20 event messages in the events list
• If the Extended HistoROM application package (order option) is enabled: up to 100
event messages are displayed in the events list along with a time stamp, plain text
description and remedial measures
• The events list can be exported and displayed via a variety of interfaces and operating
tools e.g. DeviceCare, FieldCare or Web server
Data logging
Manual
If the Extended HistoROM application package (order option) is enabled:
• Record up to 1 000 measured values via 1 to 4 channels
• User configurable recording interval
• Record up to 250 measured values via each of the 4 memory channels
• Export the measured value log via a variety of interfaces and operating tools e.g.
FieldCare, DeviceCare or web server

16.12 Certificates and approvals

Currently available certificates and approvals can be called up via the product
configurator.
The device meets the legal requirements of the applicable EU Directives. These are listed in
the corresponding EU Declaration of Conformity along with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
The devices are certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this
document on the nameplate.
• 3-A approval
• Only measuring devices with the order code for "Additional approval", option LP "3A"
have 3-A approval.
• The 3-A approval refers to the measuring device.
• When installing the measuring device, ensure that no liquid can accumulate on the
outside of the measuring device.
Remote transmitters must be installed in accordance with the 3-A Standard.
• Accessories (e.g. heating jacket, weather protection cover, wall holder unit) must be
installed in accordance with the 3-A Standard.
Each accessory can be cleaned. Disassembly may be necessary under certain
circumstances.
• EHEDG-tested
Only devices with the order code for "Additional approval", option LT "EHEDG" have been
tested and meet the requirements of the EHEDG.
To meet the requirements for EHEDG certification, the device must be used with process
connections in accordance with the EHEDG position paper entitled "Easy Cleanable Pipe
Couplings and Process Connections" (www.ehedg.org).
• FDA
• Food Contact Materials Regulation (EC) 1935/2004
Proline Promass E 500 PROFINET
Endress+Hauser

Advertisement

Table of Contents
loading

Table of Contents