Diagnostics and troubleshooting

174
12
Diagnostics and troubleshooting
12.1

General troubleshooting

For local display
Error
Local display dark and no output
signals
Local display dark and no output
signals
Local display dark and no output
signals
Local display dark and no output
signals
Local display dark and no output
signals
Local display dark and no output
signals
Local display dark and no output
signals
Local display is dark, but signal
output is within the valid range
Local display is dark, but signal
output is within the valid range
Local display is dark, but signal
output is within the valid range
Backlighting of local display is red
Text on local display appears in a
foreign language and cannot be
understood.
Message on local display:
"Communication Error"
"Check Electronics"
Possible causes
Supply voltage does not match the
value indicated on the nameplate.
The polarity of the supply voltage is
wrong.
No contact between connecting
cables and terminals.
Terminals are not plugged into the
I/O electronics module correctly.
Terminals are not plugged into the
main electronics module correctly.
I/O electronics module is defective.
Main electronics module is
defective.
The connector between the main
electronics module and display
module is not plugged in correctly.
The connecting cable is not plugged
in correctly.
Display is set too bright or too dark. • Set the display brighter by
The cable of the display module is
not plugged in correctly.
Display module is defective.
Diagnostic event with "Alarm"
diagnostic behavior has occurred.
Incorrect operating language is
configured.
Communication between the
display module and the electronics
is interrupted.
Proline Promass E 500 PROFINET
Solution
Apply the correct supply voltage .
Correct the polarity.
Check the connection of the cables
and correct if necessary.
Check terminals.
Order spare part →  248.
Check the connection and correct if
necessary.
1. Check the connection of the
electrode cable and correct if
necessary.
2. Check the connection of the coil
current cable and correct if
necessary.
simultaneously pressing  + .
• Set the display darker by
simultaneously pressing  + .
Insert the plug correctly into the
main electronics module and
display module.
Order spare part →  248.
Take remedial measures →  189
1. Press  +  for 2 s ("home
position").
2. Press .
3. Set the desired language in the
Display language parameter
(→  149).
• Check the cable and the
connector between the main
electronics module and display
module.
• Order spare part →  248.