Operating Instructions
Proline Promass E TB2
HART
Coriolis mass flow measuring system
BA01027D/06/EN/06.10
71112405
Valid as of version
01.00.zz (Device software)

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass E TB2 HART

  • Page 1 Operating Instructions Proline Promass E TB2 HART Coriolis mass flow measuring system BA01027D/06/EN/06.10 71112405 Valid as of version 01.00.zz (Device software)
  • Page 2 • The manufacturer reserves the right to modify technical data in line with technological developments without prior notice. Your Endress+Hauser sales center will supply you with current information and updates to this manual.
  • Page 3: Table Of Contents

    Proline Promass E TB2 HART Table of contents Table of contents 6.2.4 Turning the transmitter housing ..22 Important document information ..5 6.2.5 Turning the display module ... 23 About this document .
  • Page 4 10.4.1 Selecting and setting the medium ..53 14.3 Endress+Hauser services ....93 10.4.2 Configuring the current outputs ..55 10.4.3 Configuring the local display .
  • Page 5: Important Document Information

    The document types listed are available: • On the CD supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download 1.1.3 Supplementary device-dependent documentation If operating the device in SIL applications, hazardous areas or in accordance with the Pressure Equipment Directive, comply strictly with the instructions in the supplementary documentation.
  • Page 6: Document Conventions

    The Installation Instructions contain all the information needed to install the ordered accessory or spare part. The document types listed are available: • On the CD supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download Document conventions 1.2.1 Safety symbols...
  • Page 7: Symbols And Notation For Certain Types Of Information

    Proline Promass E TB2 HART Important document information Symbol Meaning Phillips head screwdriver A0011219 Allen key A0011221 Hexagon wrench A0011222 1.2.4 Symbols and notation for certain types of information Symbol Meaning Allowed Indicates procedures, processes or actions that are allowed.
  • Page 8: Basic Safety Instructions

    Observe the specified maximum process pressure. ► Verification for borderline cases: For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying ► the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability.
  • Page 9: Workplace Safety

    It fulfills general safety requirements and legal requirements. It also conforms to the EC directives listed in the device-specific EC declaration of conformity. Endress+Hauser confirms this fact by applying the CE mark. Endress+Hauser...
  • Page 10: Product Description

    Ò TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA Ò SWAGELOK Registered trademark of Swagelok & Co., Solon, USA Ò Ò FieldCare , Field Xpert , Applicator Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH Endress+Hauser...
  • Page 11: Patents

    Proline Promass E TB2 HART Product description Patents US patents EP patents 5,610,342 0 701 213 6,513,393 1 055 102 6,412,35 1 054 240 6,651,513 6,412,355 1 054 240 7,555,397 Endress+Hauser...
  • Page 12: Incoming Acceptance And Product

    DELIVERY NOTE A0013696 A0013699 Do the nameplate data match the ordering information on the delivery note? A0013697 A0013696 Is the CD-ROM with the Technical Documentation and documents present? If one of the conditions does not comply, contact your Endress+Hauser distributor. Endress+Hauser...
  • Page 13: Product Identification

    Proline Promass E TB2 HART Incoming acceptance and product identification Product identification The following options are available for identification of the measuring device: • Nameplate specifications • Order code with breakdown of the device features on the delivery note • Enter serial numbers from nameplates in W@M Device Viewer (www.endress.com/deviceviewer): All information about the measuring device is displayed.
  • Page 14 Incoming acceptance and product identification Proline Promass E TB2 HART Sensor 3 4 5 Order code: Ser. no.: Ext. ord. cd.: K-factor: Size: p test = Materials: 0044 Patents Date: 322541-0000 14 15 A0013907 å 3 Example of a sensor nameplate...
  • Page 15: Storage, Transport, Packaging

    Proline Promass E TB2 HART Storage, transport, packaging disposal Storage, transport, packaging disposal Storage and transportation conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections. They prevent mechanical damage to the sealing surfaces and fouling in the measuring tube.
  • Page 16: Disposal Instructions For Packaging

    Storage, transport, packaging disposal Proline Promass E TB2 HART Disposal instructions for packaging All packaging materials are environmentally friendly and 100% recyclable: • Measuring device secondary packaging: polymer stretch film that conforms to EC Directive 2002/95/EC (RoHS). • Packaging: – Wood crate, treated in accordance with ISPM 15 standard, which is confirmed by the affixed IPPC logo.
  • Page 17: Mounting

    Proline Promass E TB2 HART Mounting Mounting Mounting requirements No special measures such as supports are necessary. External forces are absorbed by the construction of the device. 6.1.1 Suitable mounting position Mounting location A0013499 To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: •...
  • Page 18 Mounting Proline Promass E TB2 HART A0013500 å 4 Installation in a down pipe (e.g. for batching applications) Supply tank Sensor Orifice plate, pipe restriction Valve Batching tank Æ Orifice plate, pipe restriction [mm] [in] [mm] [in] 0.24 ½ 0.40 0.55...
  • Page 19: Requirements From Environment And Process

    Proline Promass E TB2 HART Mounting Orientation Compact version Horizontal orientation, transmitter head at side A0013506 Applications with high process temperatures can increase the ambient temperature. This orientation is recommended to observe the maximum ambient temperature for the transmitter. If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to the fluid properties.
  • Page 20 Mounting Proline Promass E TB2 HART Ensure the system pressure is sufficiently high to prevent cavitation and outgassing. ► For this reason, the following mounting locations are recommended: A0013935 å 6 Downstream from pumps (no danger of vacuum) A0013936 å 7...
  • Page 21: Additional Mounting Instructions

    Proline Promass E TB2 HART Mounting Heating options If a fluid requires that no heat loss should occur at the sensor, users can avail of the following heating options: • Electrical heating, e.g. with electric band heaters • Via pipes carrying hot water or steam •...
  • Page 22: Mounting The Measuring Device

    Mounting Proline Promass E TB2 HART Mounting the measuring device 6.2.1 Required mounting tools • For flanges and other process connections: Corresponding mounting tools • For turning the transmitter housing: Open-ended wrench 8 mm (0.31 in) • For opening the securing clamps: Allen key M3 6.2.2...
  • Page 23: Turning The Display Module

    Proline Promass E TB2 HART Mounting max. 350° 8 mm 8 mm A0013713 Unscrew the securing screw using an open-ended wrench. Rotate the housing in the desired direction. Firmly tighten the securing screw. 6.2.5 Turning the display module 3 mm A0013905 Loosen the securing clamp of the electronics compartment cover using an Allen key.
  • Page 24: Post-Mounting Check

    Mounting Proline Promass E TB2 HART Post-mounting check Is the device damaged (visual inspection)? Does the measuring device conform to the measuring point specifications? For example: • Process temperature (® ä 103) • Process pressure (refer to the chapter on "Material load curves" of the "Technical Information" document) •...
  • Page 25: Electrical Connection

    Proline Promass E TB2 HART Electrical connection Electrical connection Preparing the measuring device for the electrical connection 7.1.1 Required tools • For cable entries: Use corresponding tools • For securing clamp: Allen key M3 • Wire stripper • When using stranded cables: Crimping tool for wire end ferrule •...
  • Page 26: Connecting The Signal Cables

    Electrical connection Proline Promass E TB2 HART 7.2.1 Connecting the signal cables A0013836 Loosen the securing clamp of the connection compartment cover. Unscrew the connection compartment cover. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry.
  • Page 27: Connecting To The Supply Unit

    Proline Promass E TB2 HART Electrical connection NOTICE! Housing degree of protection voided due to insufficient sealing of the housing. Screw in the thread without using lubricant. The cover threads are coated with a dry lubricant. Screw the connection compartment cover closed and tighten the securing clamp.
  • Page 28: Ensuring The Degree Of Protection Of The Measuring Device

    Electrical connection Proline Promass E TB2 HART R [ ] U [V] A0013563 Operating range Ensuring the degree of protection of the measuring device The measuring device fulfills all the requirements for the IP 66 and 67 (NEMA 4X) degree of protection.
  • Page 29 Proline Promass E TB2 HART Electrical connection If supply voltage is present, is the device ready for operation and do values appear on the display module? Are all housing covers installed and firmly tightened? Is the securing clamp tightened correctly?
  • Page 30: Operation Options

    Operation options Proline Promass E TB2 HART Operation options Overview of operation options A0013902 Local operation via display module Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX100 Field Communicator 375, 475 PLC (programmable logic controller)
  • Page 31: Structure And Function Of The Operating Menu

    Proline Promass E TB2 HART Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Display/operat. Access status display Locking status Display Operation Setup Wizard 1 Wizard 2...
  • Page 32: Operating Philosophy

    Operation options Proline Promass E TB2 HART 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles. Each user role corresponds to typical tasks within the device lifecycle. Menu User role and tasks Content/meaning...
  • Page 33: Access To The Operating Menu Via The Local Display

    Proline Promass E TB2 HART Operation options Access to the operating menu via the local display 8.3.1 Measured value display and operating elements Measured value display X X X X X X X 20.50 A0013712 Measured value display Device tag (® ä 67)
  • Page 34 Operation options Proline Promass E TB2 HART Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured value type Channel number Status symbol ¯ ¯ ¯ Example A0013962 A0013945 A0013948 Appears only if a diagnostics event is present for this process variable.
  • Page 35: Navigation View And Operating Elements

    Proline Promass E TB2 HART Operation options 8.3.2 Navigation view and operating elements In submenu In wizard /../Display/operat. 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Mass flow Locking status Display A0013993-EN A0013995-EN Navigation view Navigation path to current position...
  • Page 36 Operation options Proline Promass E TB2 HART Display area Icons for menus Display/operation Appears: • In the menu next to the "Display/operat." selection A0013973 • At the left in the navigation path in the "Display/operat." menu Setup Appears: • In the menu next to the "Setup" selection A0013974 •...
  • Page 37: Editing View And Operating Elements

    Proline Promass E TB2 HART Operation options Escape key combination (press keys simultaneously) • Pressing the key briefly – Exits the current menu level and takes you to the next higher level. – If help text is open, closes the help text of the parameter.
  • Page 38 Operation options Proline Promass E TB2 HART Text editor symbols Selection of letters from A to Z Toggle ABC _ • Between upper-case and lower- … case letters A0013981 • For entering numbers • For entering special characters A0013997 Operating symbols Confirms selection.
  • Page 39: Opening The Context Menu

    Proline Promass E TB2 HART Operation options 8.3.4 Opening the context menu Using the context menu, the user can call up the following three menus quickly and directly from the measured value display: • Setup • Conf. backup disp. • Simulation Calling up and closing the context menu The user is in the measured value display.
  • Page 40: Navigating And Selecting From List

    Operation options Proline Promass E TB2 HART 8.3.5 Navigating and selecting from list Various operating elements are used for navigation in the operating menu. While doing so, the navigation path appears at the left in the header. The individual menus are identified by symbols at the left, which are also displayed in the header during navigation.
  • Page 41: Calling Up Help Text

    Proline Promass E TB2 HART Operation options 8.3.6 Calling up help text For some parameters, help texts exist, which the user can call up from the navigation view. These briefly describe the function of the parameter and thus support fast and reliable commissioning.
  • Page 42: Changing The Parameters

    Operation options Proline Promass E TB2 HART 8.3.7 Changing the parameters For a description of the editing display - consisting of text editor and numeric editor - with symbols and operating elements (® ä 37) Example: Changing the parameter "20 mA value" to 20 kg/s Curr.
  • Page 43: User Roles And Related Access Authorization

    Proline Promass E TB2 HART Operation options A message is displayed if the value entered is outside the permitted value range. Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN 8.3.8 User roles and related access authorization The two user roles "Operator"...
  • Page 44 Operation options Proline Promass E TB2 HART By simultaneously pressing the S + O + F keys. ► Ã After enabling the keypad lock: XXXXXXXXXX 20.50 Keylock on kg/h A0014000-EN After disabling the keypad lock: XXXXXXXXXX 20.50 Keylock off kg/h...
  • Page 45: Access To The Operating Menu Via The Operating Tool

    FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and help you manage them. By using the status information, it is also a simple but effective way of checking their status and condition. Access takes place via HART protocol or CDI interface.
  • Page 46: Ams Device Manager

    Operation options Proline Promass E TB2 HART User interface A0013994-EN 8.4.3 AMS Device Manager Function scope Program from Emerson Process Management for operating and configuring measuring devices via HART protocol. Source for device description files See data (® ä 49) 8.4.4...
  • Page 47: Connecting Operating Tools

    Proline Promass E TB2 HART Operation options Source for device description files See data (® ä 49) 8.4.6 Connecting operating tools Via HART protocol A0013764 å 9 Options for remote operation via HART protocol PLC (programmable logic controller) Transmitter power supply unit, e.g. RN221N (with communication resistor)
  • Page 48 Operation options Proline Promass E TB2 HART Via CDI interface A0014019 å 10 Operating options via CDI interface Computer with operating tool (FieldCare) Commubox FXA291, connected to the CDI interface of the device Endress+Hauser...
  • Page 49: Integrating A Device Via Hart

    Proline Promass E TB2 HART Integrating a device via HART protocol Integrating a device via HART protocol Overview of device description files Version data for the device Firmware version 01.00.zz • On the title page of the Operating Instructions • On transmitter nameplate (® ä 13) •...
  • Page 50: Other Settings

    Integrating a device via HART protocol Proline Promass E TB2 HART • Expert ® Communication ® HART output ® Assign PV • Expert ® Communication ® HART output ® Assign SV • Expert ® Communication ® HART output ® Assign TV •...
  • Page 51: Commissioning

    Proline Promass E TB2 HART Commissioning Commissioning 10.1 Function check Before commissioning the measuring device, ensure that the post-installation and post-connection check have been carried out. • Checklist for "Post-mounting check" (® ä 24) • Checklist for "Post-connection check" (® ä 28) 10.2...
  • Page 52: Configuring The Measuring Device

    Commissioning Proline Promass E TB2 HART 10.4 Configuring the measuring device The Setup menu with its guided wizards contains all parameters needed for standard operation. Navigation to the "Setup" menu X X X X X X X 20.50 0104-1 Main menu...
  • Page 53: Selecting And Setting The Medium

    Proline Promass E TB2 HART Commissioning 10.4.1 Selecting and setting the medium The Select medium wizard guides you systematically through all parameters that have to be configured for selecting and setting the medium. Navigation path "Setup" menu ® Select medium...
  • Page 54 Commissioning Proline Promass E TB2 HART Parameters Description Selection/ Factory setting User entry Reference sound velocity Enter the sound velocity of 0 to 99 999 m/s 0 m/s the gas at 0°C (32°F). Temperature coefficient Enter the temperature Max. 15-digit, positive...
  • Page 55: Configuring The Current Outputs

    Proline Promass E TB2 HART Commissioning 10.4.2 Configuring the current outputs The two wizards Current output 1-2 guide you systematically through all parameters that have to be set in the configuration of the respective current output. Navigation path "Setup" menu ® Current output 1 "Setup"...
  • Page 56 Commissioning Proline Promass E TB2 HART Corrected volume Reference density flow Corrected volume Reference density flow unit unit Corrected volume flow calculation Fixed reference Calculated ref. density density Reference density Linear unit expansion coefficient Fixed reference Square density expansion coefficient...
  • Page 57 Proline Promass E TB2 HART Commissioning Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Assign current output 1-2 Select the process variable for • Off (only for current • Current output 1: mass current output. output 2) flow •...
  • Page 58 Commissioning Proline Promass E TB2 HART Parameters Description Selection/ Factory setting User entry Fixed reference density Enter the fixed value for the Max. 15-digit, positive Country-dependent: reference density. floating-point number • 1 kg/Nl • 62.4 lb/Scf Linear expansion coefficient Enter the linear, fluid-...
  • Page 59: Configuring The Local Display

    Proline Promass E TB2 HART Commissioning 10.4.3 Configuring the local display The Display wizard guides you systematically through all parameters that can configured for configuring the local display. Navigation path "Setup" menu ® Display Sequence of the wizard Format display...
  • Page 60 Commissioning Proline Promass E TB2 HART Parameters Description Selection/ Factory setting User entry Value 1 display Select the measured value • None Mass flow shown on the local display. • Mass flow • Volume flow • Corrected volume flow • Density •...
  • Page 61: Configuring The Output Conditioning

    Proline Promass E TB2 HART Commissioning Parameters Description Selection/ Factory setting User entry Value 4 display Select the measured value • None None shown on the local display. • Mass flow • Volume flow • Corrected volume flow • Density •...
  • Page 62 Commissioning Proline Promass E TB2 HART Sequence of the wizard Level of flow fluctuation Slight Moderate Strong Display Display Display damping damping damping Damping Damping Damping output 1 output 1 output 1 Damping Damping Damping output 2 output 2 output 2...
  • Page 63 Proline Promass E TB2 HART Commissioning Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Level of flow fluctuation Specify the extent measured • Slight Moderate values fluctuate due to • Moderate changes in the process. • Strong...
  • Page 64: Configuring The Low Flow Cut Off

    Commissioning Proline Promass E TB2 HART 10.4.5 Configuring the low flow cut off The Low flow cut off wizard guides you systematically through all parameters that have to be set for configuring the low flow cut off. Navigation path "Setup" menu ® Low flow cut off...
  • Page 65: Configuring The Partial Filled Pipe Detection

    Proline Promass E TB2 HART Commissioning 10.4.6 Configuring the partial filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation path "Setup"...
  • Page 66: Advanced Settings

    Commissioning Proline Promass E TB2 HART 10.5 Advanced settings The Advanced setup menu with its submenus contains all parameters needed for specific settings. Navigation path "Setup" menu ® Advanced setup Navigation to the "Advanced setup" menu X X X X X X X 20.50...
  • Page 67: Defining The Tag Name

    Proline Promass E TB2 HART Commissioning Conf. backup disp. ® (® ä 72) 10.5.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Tag name parameter and thus change the factory setting.
  • Page 68 Commissioning Proline Promass E TB2 HART Pressure unit Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Mass flow unit Select the unit for mass flow. Unit picklist Country-dependent: • kg/h Result • lb/min The selected unit applies for: –...
  • Page 69: Carrying Out A Sensor Adjustment

    Proline Promass E TB2 HART Commissioning Parameters Description Selection/ Factory setting User entry Reference density unit Select the unit for reference Unit picklist Country-dependent: density. • kg/Nl • lb/scf Result The selected unit applies for: – Current outputs – Low value partial filled pipe detection –...
  • Page 70: Configuring The Totalizer

    Commissioning Proline Promass E TB2 HART 10.5.4 Configuring the totalizer You can configure each totalizer in the three submenus Totalizer 1-3. Navigation path "Setup" menu ® Advanced setup ® Totalizer 1 "Setup" menu ® Advanced setup ® Totalizer 2 "Setup" menu ® Advanced setup ® Totalizer 3...
  • Page 71: Carry Out Additional Display Configurations

    Proline Promass E TB2 HART Commissioning 10.5.5 Carry out additional display configurations In the Display submenu, you can set all parameters involved in the configuration of the local display. Navigation path "Setup" menu ® Advanced setup ® Display Structure of the submenu Display ®...
  • Page 72: Configuration Management

    Commissioning Proline Promass E TB2 HART Parameters Prerequisite Description Selection/ Factory setting User entry Decimal places 2 A measured value is Select the number of • x x.xx specified in the digits after the decimal • x.x parameter Value 2 point for the display •...
  • Page 73: Simulation

    Proline Promass E TB2 HART Commissioning While this action is in progress, the configuration cannot be edited via the local display and a message on the processing status appears on the display. 10.7 Simulation The Simulation submenu enables you to simulate, without a real flow situation, various process variables in the process and the device alarm mode and to verify downstream signal chains (switching valves or closed-control loops).
  • Page 74: Protecting Settings From Unauthorized Access

    Commissioning Proline Promass E TB2 HART Parameters Prerequisite Description Selection/ Factory setting User entry Simulation device Switch the device • On alarm alarm on and off. • Off 10.8 Protecting settings from unauthorized access The following options exist for protecting the configuration of the measuring device from unintentional modification after commissioning: •...
  • Page 75: Write Protection Via Lock Switch

    Proline Promass E TB2 HART Commissioning 10.8.2 Write protection via lock switch Unlike write protection via user-specific access code, this allows write access to the entire operating menu - other than the Contrast display parameter - to be locked. The values of the parameters are still visible, but can no longer be changed (except for Contrast display), either via the local display, CDI interface or HART protocol.
  • Page 76 Commissioning Proline Promass E TB2 HART A0013702 If the hardware write protection is disabled, the -symbol disappears in the header of the measured value display and in the navigation view in front of the parameters. Feed the spiral cable into the gap between the housing and main electronics module and plug the display module into the electronics compartment in the desired direction until it engages.
  • Page 77: Operation

    Proline Promass E TB2 HART Operation Operation 11.1 Adjusting the operating language Information (® ä 51) 11.2 Configuring the display • Basic settings for local display (® ä 59) • Advanced settings for local display (® ä 71) 11.3 Reading off measured values You can read all measured values using the Measured value menu.
  • Page 78: Adapting The Measuring Device To The Process Conditions

    Operation Proline Promass E TB2 HART Output current 1 Measured current 1 Terminal voltage 1 Output current 2 11.4 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using the Setup menu (® ä 52) •...
  • Page 79: Show Data Logging

    Proline Promass E TB2 HART Operation Function scope of parameter "Reset all totalizers" Description Resets all the totalizers to 0 and restarts the totaling process. This deletes all the flow values previously totalized. 11.6 Show data logging In the device, the extended function of the HistoROM must be enabled (order option) so that the measured value memory submenu appears.
  • Page 80 Operation Proline Promass E TB2 HART If the length of the logging interval or the assignment of the process variables to the channels is changed, the content of the data logging is deleted. Endress+Hauser...
  • Page 81: Diagnostics And Troubleshooting

    Proline Promass E TB2 HART Diagnostics and troubleshooting Diagnostics and troubleshooting 12.1 General troubleshooting Problem Possible cause Remedy Local display dark and no signal Supply voltage does not match that Apply the correct supply voltage output at current output (0 mA) specified on the nameplate.
  • Page 82: Diagnostics Event On Local Display

    Diagnostics and troubleshooting Proline Promass E TB2 HART 12.2 Diagnostics event on local display 12.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the measured value display.
  • Page 83: Calling Up Remedial Measures

    Proline Promass E TB2 HART Diagnostics and troubleshooting Diagnostics event and event text The fault can be identified using the diagnostics event. The event text helps you by providing information about the fault. In addition, the corresponding symbol is displayed before the diagnostics event.
  • Page 84: Diagnostics Event In The Operating Tool

    Diagnostics and troubleshooting Proline Promass E TB2 HART Ã The message for the remedial measures for the diagnostics event opens. Press S + O simultaneously. Ã The message about the remedial measures closes. 12.3 Diagnostics event in the operating tool...
  • Page 85: Overview Of Diagnostics Events

    Proline Promass E TB2 HART Diagnostics and troubleshooting / ../Diagnose list Diagnostics F273 Main electronic Diagnostics 2 Diagnostics 3 A0014006-EN Calling up and closing the remedial measures Press F. Ã The message for the remedial measures for the selected diagnostics event opens.
  • Page 86: Diagnostics For The Electronics

    Diagnostics and troubleshooting Proline Promass E TB2 HART Event level from the Diagnostics event Event text Remedial measures factory S044 Sensor drift Electrodynamic sensor signal deviation! Warning 1. Check or change main electronics. 2. Change sensor. S046 Sensor limit The measuring tube oscillation frequency is...
  • Page 87: Diagnostics For The Configuration

    Proline Promass E TB2 HART Diagnostics and troubleshooting 12.5.4 Diagnostics for the configuration Event level from the Diagnostics event Event text Remedial measures factory F410 Data transfer 1. Check connection. Alarm 2. Retry data transfer. F411 Up-/Download 1. Check connection.
  • Page 88: Event Logbook

    Diagnostics and troubleshooting Proline Promass E TB2 HART Event level from the Diagnostics event Event text Remedial measures factory S912 Inhomogeneous Fluid is inhomogeneous, e.g. gas or solid Warning content! 1. Check process conditions. 2. Increase system pressure. In particular with out gassing fluids and/...
  • Page 89: Filtering The Event Logbook

    Proline Promass E TB2 HART Diagnostics and troubleshooting Calling up and closing the remedial measures Press F. Ã The message for the remedial measures for the selected diagnostics event opens. Press S + O simultaneously. Ã The message about the remedial measures closes.
  • Page 90 Diagnostics and troubleshooting Proline Promass E TB2 HART Information Event text event I1221 Error during zero point adjustment I1222 Zero point adjustment OK I1227 Sensor emergency mode enabled I1228 Sensor emergency mode failed Endress+Hauser...
  • Page 91: Repair

    • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices by Endress+Hauser Service or at the factory only.
  • Page 92: Endress+Hauser Services

    Repair Proline Promass E TB2 HART 13.3 Endress+Hauser services For information about service and spare parts, contact your Endress+Hauser distributor. Endress+Hauser...
  • Page 93: Maintenance

    • Observe the maximum permitted fluid temperature for the measuring device (® ä 103). 14.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser representative can provide detailed information on the services.
  • Page 94: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 95: Return

    • Complete all fields of the "Declaration of Contamination" form and enclose it with the returned device. Only then can Endress+Hauser transport, examine and repair a returned device. • If necessary, enclose special handling regulations, such as a safety data sheet according to (EC) regulation No.
  • Page 96: Disposal

    Disposal Proline Promass E TB2 HART Disposal 17.1 Removing the measuring device Switch off the device. WARNING! Danger to persons from process conditions. Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
  • Page 97: Technical Data

    Proline Promass E TB2 HART Technical data Technical data 18.1 Application The measuring device is suitable for flow measurement of liquids and gases only. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing fluids.
  • Page 98: Output

    Technical data Proline Promass E TB2 HART Maximum full scale value for liquid [kg/h] max(F) < g can never be greater than g max(G) max(F) max(G) max(F) Gas density in [kg/m ] at operating conditions [mm] [in] [kg/m ½ 1 ½...
  • Page 99: Power Supply

    Proline Promass E TB2 HART Technical data Protocol-specific data • For information on the device description files (® ä 49) • For information on the HART device variables and measured values • For information on the HART commands 18.5 Power supply...
  • Page 100: Performance Characteristics

    Technical data Proline Promass E TB2 HART 18.6 Performance characteristics Reference operating • Error limits following ISO/DIS 11631 conditions • Water with +20 to +30 °C (+68 to +86 °F) at 2 to 4 bar (29 to 58 psi) • Specifications as per calibration protocol ±5 °C (±9 °F) and ±2 bar (±29 psi) •...
  • Page 101 Proline Promass E TB2 HART Technical data Example for max. measured error ±1.0 ±0.5 ±0.2 18 t/h A0010443 å 20 Maximum measured error in % o.r. (example: Promass E TB2, DN 25) Design fundamentals (® ä 102) Flow values (example)
  • Page 102: Operating Conditions: Installation

    Technical data Proline Promass E TB2 HART Influence of ambient Current output (additional error, in relation to the span of 16 mA): temperature – Zero point (4 mA) Average T : 0.02 %/10 K, max. 0.35 % over the entire temperature range –40 to +60 °C (–40 to +140 °F)
  • Page 103: Operating Conditions: Process

    Proline Promass E TB2 HART Technical data If operating outdoors: ► Avoid direct sunlight, particularly in warm climatic regions. Storage temperature –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) Climate class DIN EN 60068-2-38 (test Z/AD)
  • Page 104 Technical data Proline Promass E TB2 HART • The minimum recommended full scale value is approx. 1/20 of the maximum full scale value • In most applications, 20 to 50% of the maximum full scale value can be considered ideal •...
  • Page 105: 18.10 Mechanical Construction

    Proline Promass E TB2 HART Technical data [mbar] 10000 DN 40 DN 25 DN 15 DN 50 DN 8 1000 0.01 0.001 1000 [t/h] A0013674 å 21 Pressure loss diagram for water Pressure loss in US units Pressure loss depends on the fluid properties and the nominal diameter.
  • Page 106 Technical data Proline Promass E TB2 HART Weight [lb] [in] With transmitter housing GT20, With transmitter housing GT18 aluminum coating 1.4404/316L 13.2 18.7 ½ 14.3 19.8 17.6 23.2 1 ½ 28.7 34.2 48.5 54.0 Materials Transmitter housing • Compact version –...
  • Page 107: 18.11 Human Interface

    Proline Promass E TB2 HART Technical data With transmitter housing GT18 1.4404/316L Electrical connection Type of protection Material Cable gland M20 × 1.5 • Non-Ex Stainless steel 1.4404 • Ex ia • Ex ic • Ex nA • Ex t Thread G ½"...
  • Page 108: 18.12 Certificates And Approvals

    Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 109: 18.13 Application Packages

    • NAMUR NE 132 Coriolis mass meter Pressure Equipment Directive • With the PED/G1/III marking on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure Equipment Directive 97/23/EC. • Devices not bearing this marking (PED) are designed and manufactured according to good engineering practice.
  • Page 110: Index

    Index Proline Promass E TB2 HART Index Symbols Connection tools ......25...
  • Page 111 Proline Promass E TB2 HART Index Commubox FXA191, 195 ....47 FieldCare ....... . 45 Commubox FXA291 .
  • Page 112 Index Proline Promass E TB2 HART Local display Mounting requirements see Measured value display Inlet and outlet runs ......19 see Diagnostics message Installation dimensions .
  • Page 113 Proline Promass E TB2 HART Index For the local display ..... . 59, 71 Sensor For totalizer ......70 Medium temperature range .
  • Page 114 Index Proline Promass E TB2 HART Supply voltage (Power supply) ....27, 99 Symbols Version data for the device ..... . 49 For channel number .
  • Page 115 Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
  • Page 116 www.endress.com/worldwide *71112405* BA01027D/06/EN/06.10 71112405 EH-COSIMA ProMoDo 71112405...

Table of Contents