Summary of Contents for Endress+Hauser Proline Promass E TB2 HART
Page 1
Operating Instructions Proline Promass E TB2 HART Coriolis mass flow measuring system BA01027D/06/EN/06.10 71112405 Valid as of version 01.00.zz (Device software)
Page 2
• The manufacturer reserves the right to modify technical data in line with technological developments without prior notice. Your Endress+Hauser sales center will supply you with current information and updates to this manual.
Proline Promass E TB2 HART Table of contents Table of contents 6.2.4 Turning the transmitter housing ..22 Important document information ..5 6.2.5 Turning the display module ... 23 About this document .
Page 4
10.4.1 Selecting and setting the medium ..53 14.3 Endress+Hauser services ....93 10.4.2 Configuring the current outputs ..55 10.4.3 Configuring the local display .
The document types listed are available: • On the CD supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download 1.1.3 Supplementary device-dependent documentation If operating the device in SIL applications, hazardous areas or in accordance with the Pressure Equipment Directive, comply strictly with the instructions in the supplementary documentation.
The Installation Instructions contain all the information needed to install the ordered accessory or spare part. The document types listed are available: • On the CD supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download Document conventions 1.2.1 Safety symbols...
Proline Promass E TB2 HART Important document information Symbol Meaning Phillips head screwdriver A0011219 Allen key A0011221 Hexagon wrench A0011222 1.2.4 Symbols and notation for certain types of information Symbol Meaning Allowed Indicates procedures, processes or actions that are allowed.
Observe the specified maximum process pressure. ► Verification for borderline cases: For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying ► the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability.
It fulfills general safety requirements and legal requirements. It also conforms to the EC directives listed in the device-specific EC declaration of conformity. Endress+Hauser confirms this fact by applying the CE mark. Endress+Hauser...
Ò TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA Ò SWAGELOK Registered trademark of Swagelok & Co., Solon, USA Ò Ò FieldCare , Field Xpert , Applicator Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH Endress+Hauser...
DELIVERY NOTE A0013696 A0013699 Do the nameplate data match the ordering information on the delivery note? A0013697 A0013696 Is the CD-ROM with the Technical Documentation and documents present? If one of the conditions does not comply, contact your Endress+Hauser distributor. Endress+Hauser...
Proline Promass E TB2 HART Incoming acceptance and product identification Product identification The following options are available for identification of the measuring device: • Nameplate specifications • Order code with breakdown of the device features on the delivery note • Enter serial numbers from nameplates in W@M Device Viewer (www.endress.com/deviceviewer): All information about the measuring device is displayed.
Page 14
Incoming acceptance and product identification Proline Promass E TB2 HART Sensor 3 4 5 Order code: Ser. no.: Ext. ord. cd.: K-factor: Size: p test = Materials: 0044 Patents Date: 322541-0000 14 15 A0013907 å 3 Example of a sensor nameplate...
Proline Promass E TB2 HART Storage, transport, packaging disposal Storage, transport, packaging disposal Storage and transportation conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections. They prevent mechanical damage to the sealing surfaces and fouling in the measuring tube.
Storage, transport, packaging disposal Proline Promass E TB2 HART Disposal instructions for packaging All packaging materials are environmentally friendly and 100% recyclable: • Measuring device secondary packaging: polymer stretch film that conforms to EC Directive 2002/95/EC (RoHS). • Packaging: – Wood crate, treated in accordance with ISPM 15 standard, which is confirmed by the affixed IPPC logo.
Proline Promass E TB2 HART Mounting Mounting Mounting requirements No special measures such as supports are necessary. External forces are absorbed by the construction of the device. 6.1.1 Suitable mounting position Mounting location A0013499 To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: •...
Page 18
Mounting Proline Promass E TB2 HART A0013500 å 4 Installation in a down pipe (e.g. for batching applications) Supply tank Sensor Orifice plate, pipe restriction Valve Batching tank Æ Orifice plate, pipe restriction [mm] [in] [mm] [in] 0.24 ½ 0.40 0.55...
Proline Promass E TB2 HART Mounting Orientation Compact version Horizontal orientation, transmitter head at side A0013506 Applications with high process temperatures can increase the ambient temperature. This orientation is recommended to observe the maximum ambient temperature for the transmitter. If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to the fluid properties.
Page 20
Mounting Proline Promass E TB2 HART Ensure the system pressure is sufficiently high to prevent cavitation and outgassing. ► For this reason, the following mounting locations are recommended: A0013935 å 6 Downstream from pumps (no danger of vacuum) A0013936 å 7...
Proline Promass E TB2 HART Mounting Heating options If a fluid requires that no heat loss should occur at the sensor, users can avail of the following heating options: • Electrical heating, e.g. with electric band heaters • Via pipes carrying hot water or steam •...
Mounting Proline Promass E TB2 HART Mounting the measuring device 6.2.1 Required mounting tools • For flanges and other process connections: Corresponding mounting tools • For turning the transmitter housing: Open-ended wrench 8 mm (0.31 in) • For opening the securing clamps: Allen key M3 6.2.2...
Proline Promass E TB2 HART Mounting max. 350° 8 mm 8 mm A0013713 Unscrew the securing screw using an open-ended wrench. Rotate the housing in the desired direction. Firmly tighten the securing screw. 6.2.5 Turning the display module 3 mm A0013905 Loosen the securing clamp of the electronics compartment cover using an Allen key.
Mounting Proline Promass E TB2 HART Post-mounting check Is the device damaged (visual inspection)? Does the measuring device conform to the measuring point specifications? For example: • Process temperature (® ä 103) • Process pressure (refer to the chapter on "Material load curves" of the "Technical Information" document) •...
Proline Promass E TB2 HART Electrical connection Electrical connection Preparing the measuring device for the electrical connection 7.1.1 Required tools • For cable entries: Use corresponding tools • For securing clamp: Allen key M3 • Wire stripper • When using stranded cables: Crimping tool for wire end ferrule •...
Electrical connection Proline Promass E TB2 HART 7.2.1 Connecting the signal cables A0013836 Loosen the securing clamp of the connection compartment cover. Unscrew the connection compartment cover. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry.
Proline Promass E TB2 HART Electrical connection NOTICE! Housing degree of protection voided due to insufficient sealing of the housing. Screw in the thread without using lubricant. The cover threads are coated with a dry lubricant. Screw the connection compartment cover closed and tighten the securing clamp.
Electrical connection Proline Promass E TB2 HART R [ ] U [V] A0013563 Operating range Ensuring the degree of protection of the measuring device The measuring device fulfills all the requirements for the IP 66 and 67 (NEMA 4X) degree of protection.
Page 29
Proline Promass E TB2 HART Electrical connection If supply voltage is present, is the device ready for operation and do values appear on the display module? Are all housing covers installed and firmly tightened? Is the securing clamp tightened correctly?
Proline Promass E TB2 HART Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Display/operat. Access status display Locking status Display Operation Setup Wizard 1 Wizard 2...
Operation options Proline Promass E TB2 HART 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles. Each user role corresponds to typical tasks within the device lifecycle. Menu User role and tasks Content/meaning...
Proline Promass E TB2 HART Operation options Access to the operating menu via the local display 8.3.1 Measured value display and operating elements Measured value display X X X X X X X 20.50 A0013712 Measured value display Device tag (® ä 67)
Page 34
Operation options Proline Promass E TB2 HART Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured value type Channel number Status symbol ¯ ¯ ¯ Example A0013962 A0013945 A0013948 Appears only if a diagnostics event is present for this process variable.
Proline Promass E TB2 HART Operation options 8.3.2 Navigation view and operating elements In submenu In wizard /../Display/operat. 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Mass flow Locking status Display A0013993-EN A0013995-EN Navigation view Navigation path to current position...
Page 36
Operation options Proline Promass E TB2 HART Display area Icons for menus Display/operation Appears: • In the menu next to the "Display/operat." selection A0013973 • At the left in the navigation path in the "Display/operat." menu Setup Appears: • In the menu next to the "Setup" selection A0013974 •...
Proline Promass E TB2 HART Operation options Escape key combination (press keys simultaneously) • Pressing the key briefly – Exits the current menu level and takes you to the next higher level. – If help text is open, closes the help text of the parameter.
Page 38
Operation options Proline Promass E TB2 HART Text editor symbols Selection of letters from A to Z Toggle ABC _ • Between upper-case and lower- … case letters A0013981 • For entering numbers • For entering special characters A0013997 Operating symbols Confirms selection.
Proline Promass E TB2 HART Operation options 8.3.4 Opening the context menu Using the context menu, the user can call up the following three menus quickly and directly from the measured value display: • Setup • Conf. backup disp. • Simulation Calling up and closing the context menu The user is in the measured value display.
Operation options Proline Promass E TB2 HART 8.3.5 Navigating and selecting from list Various operating elements are used for navigation in the operating menu. While doing so, the navigation path appears at the left in the header. The individual menus are identified by symbols at the left, which are also displayed in the header during navigation.
Proline Promass E TB2 HART Operation options 8.3.6 Calling up help text For some parameters, help texts exist, which the user can call up from the navigation view. These briefly describe the function of the parameter and thus support fast and reliable commissioning.
Operation options Proline Promass E TB2 HART 8.3.7 Changing the parameters For a description of the editing display - consisting of text editor and numeric editor - with symbols and operating elements (® ä 37) Example: Changing the parameter "20 mA value" to 20 kg/s Curr.
Proline Promass E TB2 HART Operation options A message is displayed if the value entered is outside the permitted value range. Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN 8.3.8 User roles and related access authorization The two user roles "Operator"...
Page 44
Operation options Proline Promass E TB2 HART By simultaneously pressing the S + O + F keys. ► Ã After enabling the keypad lock: XXXXXXXXXX 20.50 Keylock on kg/h A0014000-EN After disabling the keypad lock: XXXXXXXXXX 20.50 Keylock off kg/h...
FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and help you manage them. By using the status information, it is also a simple but effective way of checking their status and condition. Access takes place via HART protocol or CDI interface.
Operation options Proline Promass E TB2 HART User interface A0013994-EN 8.4.3 AMS Device Manager Function scope Program from Emerson Process Management for operating and configuring measuring devices via HART protocol. Source for device description files See data (® ä 49) 8.4.4...
Proline Promass E TB2 HART Operation options Source for device description files See data (® ä 49) 8.4.6 Connecting operating tools Via HART protocol A0013764 å 9 Options for remote operation via HART protocol PLC (programmable logic controller) Transmitter power supply unit, e.g. RN221N (with communication resistor)
Page 48
Operation options Proline Promass E TB2 HART Via CDI interface A0014019 å 10 Operating options via CDI interface Computer with operating tool (FieldCare) Commubox FXA291, connected to the CDI interface of the device Endress+Hauser...
Proline Promass E TB2 HART Integrating a device via HART protocol Integrating a device via HART protocol Overview of device description files Version data for the device Firmware version 01.00.zz • On the title page of the Operating Instructions • On transmitter nameplate (® ä 13) •...
Proline Promass E TB2 HART Commissioning Commissioning 10.1 Function check Before commissioning the measuring device, ensure that the post-installation and post-connection check have been carried out. • Checklist for "Post-mounting check" (® ä 24) • Checklist for "Post-connection check" (® ä 28) 10.2...
Commissioning Proline Promass E TB2 HART 10.4 Configuring the measuring device The Setup menu with its guided wizards contains all parameters needed for standard operation. Navigation to the "Setup" menu X X X X X X X 20.50 0104-1 Main menu...
Proline Promass E TB2 HART Commissioning 10.4.1 Selecting and setting the medium The Select medium wizard guides you systematically through all parameters that have to be configured for selecting and setting the medium. Navigation path "Setup" menu ® Select medium...
Page 54
Commissioning Proline Promass E TB2 HART Parameters Description Selection/ Factory setting User entry Reference sound velocity Enter the sound velocity of 0 to 99 999 m/s 0 m/s the gas at 0°C (32°F). Temperature coefficient Enter the temperature Max. 15-digit, positive...
Proline Promass E TB2 HART Commissioning 10.4.2 Configuring the current outputs The two wizards Current output 1-2 guide you systematically through all parameters that have to be set in the configuration of the respective current output. Navigation path "Setup" menu ® Current output 1 "Setup"...
Page 56
Commissioning Proline Promass E TB2 HART Corrected volume Reference density flow Corrected volume Reference density flow unit unit Corrected volume flow calculation Fixed reference Calculated ref. density density Reference density Linear unit expansion coefficient Fixed reference Square density expansion coefficient...
Page 57
Proline Promass E TB2 HART Commissioning Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Assign current output 1-2 Select the process variable for • Off (only for current • Current output 1: mass current output. output 2) flow •...
Page 58
Commissioning Proline Promass E TB2 HART Parameters Description Selection/ Factory setting User entry Fixed reference density Enter the fixed value for the Max. 15-digit, positive Country-dependent: reference density. floating-point number • 1 kg/Nl • 62.4 lb/Scf Linear expansion coefficient Enter the linear, fluid-...
Proline Promass E TB2 HART Commissioning 10.4.3 Configuring the local display The Display wizard guides you systematically through all parameters that can configured for configuring the local display. Navigation path "Setup" menu ® Display Sequence of the wizard Format display...
Page 60
Commissioning Proline Promass E TB2 HART Parameters Description Selection/ Factory setting User entry Value 1 display Select the measured value • None Mass flow shown on the local display. • Mass flow • Volume flow • Corrected volume flow • Density •...
Proline Promass E TB2 HART Commissioning Parameters Description Selection/ Factory setting User entry Value 4 display Select the measured value • None None shown on the local display. • Mass flow • Volume flow • Corrected volume flow • Density •...
Page 63
Proline Promass E TB2 HART Commissioning Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Level of flow fluctuation Specify the extent measured • Slight Moderate values fluctuate due to • Moderate changes in the process. • Strong...
Commissioning Proline Promass E TB2 HART 10.4.5 Configuring the low flow cut off The Low flow cut off wizard guides you systematically through all parameters that have to be set for configuring the low flow cut off. Navigation path "Setup" menu ® Low flow cut off...
Proline Promass E TB2 HART Commissioning 10.4.6 Configuring the partial filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation path "Setup"...
Commissioning Proline Promass E TB2 HART 10.5 Advanced settings The Advanced setup menu with its submenus contains all parameters needed for specific settings. Navigation path "Setup" menu ® Advanced setup Navigation to the "Advanced setup" menu X X X X X X X 20.50...
Proline Promass E TB2 HART Commissioning Conf. backup disp. ® (® ä 72) 10.5.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Tag name parameter and thus change the factory setting.
Page 68
Commissioning Proline Promass E TB2 HART Pressure unit Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Mass flow unit Select the unit for mass flow. Unit picklist Country-dependent: • kg/h Result • lb/min The selected unit applies for: –...
Proline Promass E TB2 HART Commissioning Parameters Description Selection/ Factory setting User entry Reference density unit Select the unit for reference Unit picklist Country-dependent: density. • kg/Nl • lb/scf Result The selected unit applies for: – Current outputs – Low value partial filled pipe detection –...
Commissioning Proline Promass E TB2 HART 10.5.4 Configuring the totalizer You can configure each totalizer in the three submenus Totalizer 1-3. Navigation path "Setup" menu ® Advanced setup ® Totalizer 1 "Setup" menu ® Advanced setup ® Totalizer 2 "Setup" menu ® Advanced setup ® Totalizer 3...
Proline Promass E TB2 HART Commissioning 10.5.5 Carry out additional display configurations In the Display submenu, you can set all parameters involved in the configuration of the local display. Navigation path "Setup" menu ® Advanced setup ® Display Structure of the submenu Display ®...
Commissioning Proline Promass E TB2 HART Parameters Prerequisite Description Selection/ Factory setting User entry Decimal places 2 A measured value is Select the number of • x x.xx specified in the digits after the decimal • x.x parameter Value 2 point for the display •...
Proline Promass E TB2 HART Commissioning While this action is in progress, the configuration cannot be edited via the local display and a message on the processing status appears on the display. 10.7 Simulation The Simulation submenu enables you to simulate, without a real flow situation, various process variables in the process and the device alarm mode and to verify downstream signal chains (switching valves or closed-control loops).
Commissioning Proline Promass E TB2 HART Parameters Prerequisite Description Selection/ Factory setting User entry Simulation device Switch the device • On alarm alarm on and off. • Off 10.8 Protecting settings from unauthorized access The following options exist for protecting the configuration of the measuring device from unintentional modification after commissioning: •...
Proline Promass E TB2 HART Commissioning 10.8.2 Write protection via lock switch Unlike write protection via user-specific access code, this allows write access to the entire operating menu - other than the Contrast display parameter - to be locked. The values of the parameters are still visible, but can no longer be changed (except for Contrast display), either via the local display, CDI interface or HART protocol.
Page 76
Commissioning Proline Promass E TB2 HART A0013702 If the hardware write protection is disabled, the -symbol disappears in the header of the measured value display and in the navigation view in front of the parameters. Feed the spiral cable into the gap between the housing and main electronics module and plug the display module into the electronics compartment in the desired direction until it engages.
Proline Promass E TB2 HART Operation Operation 11.1 Adjusting the operating language Information (® ä 51) 11.2 Configuring the display • Basic settings for local display (® ä 59) • Advanced settings for local display (® ä 71) 11.3 Reading off measured values You can read all measured values using the Measured value menu.
Operation Proline Promass E TB2 HART Output current 1 Measured current 1 Terminal voltage 1 Output current 2 11.4 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using the Setup menu (® ä 52) •...
Proline Promass E TB2 HART Operation Function scope of parameter "Reset all totalizers" Description Resets all the totalizers to 0 and restarts the totaling process. This deletes all the flow values previously totalized. 11.6 Show data logging In the device, the extended function of the HistoROM must be enabled (order option) so that the measured value memory submenu appears.
Page 80
Operation Proline Promass E TB2 HART If the length of the logging interval or the assignment of the process variables to the channels is changed, the content of the data logging is deleted. Endress+Hauser...
Proline Promass E TB2 HART Diagnostics and troubleshooting Diagnostics and troubleshooting 12.1 General troubleshooting Problem Possible cause Remedy Local display dark and no signal Supply voltage does not match that Apply the correct supply voltage output at current output (0 mA) specified on the nameplate.
Diagnostics and troubleshooting Proline Promass E TB2 HART 12.2 Diagnostics event on local display 12.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the measured value display.
Proline Promass E TB2 HART Diagnostics and troubleshooting Diagnostics event and event text The fault can be identified using the diagnostics event. The event text helps you by providing information about the fault. In addition, the corresponding symbol is displayed before the diagnostics event.
Diagnostics and troubleshooting Proline Promass E TB2 HART Ã The message for the remedial measures for the diagnostics event opens. Press S + O simultaneously. Ã The message about the remedial measures closes. 12.3 Diagnostics event in the operating tool...
Proline Promass E TB2 HART Diagnostics and troubleshooting / ../Diagnose list Diagnostics F273 Main electronic Diagnostics 2 Diagnostics 3 A0014006-EN Calling up and closing the remedial measures Press F. Ã The message for the remedial measures for the selected diagnostics event opens.
Diagnostics and troubleshooting Proline Promass E TB2 HART Event level from the Diagnostics event Event text Remedial measures factory S044 Sensor drift Electrodynamic sensor signal deviation! Warning 1. Check or change main electronics. 2. Change sensor. S046 Sensor limit The measuring tube oscillation frequency is...
Proline Promass E TB2 HART Diagnostics and troubleshooting 12.5.4 Diagnostics for the configuration Event level from the Diagnostics event Event text Remedial measures factory F410 Data transfer 1. Check connection. Alarm 2. Retry data transfer. F411 Up-/Download 1. Check connection.
Diagnostics and troubleshooting Proline Promass E TB2 HART Event level from the Diagnostics event Event text Remedial measures factory S912 Inhomogeneous Fluid is inhomogeneous, e.g. gas or solid Warning content! 1. Check process conditions. 2. Increase system pressure. In particular with out gassing fluids and/...
Proline Promass E TB2 HART Diagnostics and troubleshooting Calling up and closing the remedial measures Press F. Ã The message for the remedial measures for the selected diagnostics event opens. Press S + O simultaneously. Ã The message about the remedial measures closes.
Page 90
Diagnostics and troubleshooting Proline Promass E TB2 HART Information Event text event I1221 Error during zero point adjustment I1222 Zero point adjustment OK I1227 Sensor emergency mode enabled I1228 Sensor emergency mode failed Endress+Hauser...
• Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices by Endress+Hauser Service or at the factory only.
Repair Proline Promass E TB2 HART 13.3 Endress+Hauser services For information about service and spare parts, contact your Endress+Hauser distributor. Endress+Hauser...
• Observe the maximum permitted fluid temperature for the measuring device (® ä 103). 14.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser representative can provide detailed information on the services.
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
• Complete all fields of the "Declaration of Contamination" form and enclose it with the returned device. Only then can Endress+Hauser transport, examine and repair a returned device. • If necessary, enclose special handling regulations, such as a safety data sheet according to (EC) regulation No.
Disposal Proline Promass E TB2 HART Disposal 17.1 Removing the measuring device Switch off the device. WARNING! Danger to persons from process conditions. Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
Proline Promass E TB2 HART Technical data Technical data 18.1 Application The measuring device is suitable for flow measurement of liquids and gases only. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing fluids.
Technical data Proline Promass E TB2 HART Maximum full scale value for liquid [kg/h] max(F) < g can never be greater than g max(G) max(F) max(G) max(F) Gas density in [kg/m ] at operating conditions [mm] [in] [kg/m ½ 1 ½...
Proline Promass E TB2 HART Technical data Protocol-specific data • For information on the device description files (® ä 49) • For information on the HART device variables and measured values • For information on the HART commands 18.5 Power supply...
Technical data Proline Promass E TB2 HART 18.6 Performance characteristics Reference operating • Error limits following ISO/DIS 11631 conditions • Water with +20 to +30 °C (+68 to +86 °F) at 2 to 4 bar (29 to 58 psi) • Specifications as per calibration protocol ±5 °C (±9 °F) and ±2 bar (±29 psi) •...
Page 101
Proline Promass E TB2 HART Technical data Example for max. measured error ±1.0 ±0.5 ±0.2 18 t/h A0010443 å 20 Maximum measured error in % o.r. (example: Promass E TB2, DN 25) Design fundamentals (® ä 102) Flow values (example)
Technical data Proline Promass E TB2 HART Influence of ambient Current output (additional error, in relation to the span of 16 mA): temperature – Zero point (4 mA) Average T : 0.02 %/10 K, max. 0.35 % over the entire temperature range –40 to +60 °C (–40 to +140 °F)
Proline Promass E TB2 HART Technical data If operating outdoors: ► Avoid direct sunlight, particularly in warm climatic regions. Storage temperature –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) Climate class DIN EN 60068-2-38 (test Z/AD)
Page 104
Technical data Proline Promass E TB2 HART • The minimum recommended full scale value is approx. 1/20 of the maximum full scale value • In most applications, 20 to 50% of the maximum full scale value can be considered ideal •...
Proline Promass E TB2 HART Technical data [mbar] 10000 DN 40 DN 25 DN 15 DN 50 DN 8 1000 0.01 0.001 1000 [t/h] A0013674 å 21 Pressure loss diagram for water Pressure loss in US units Pressure loss depends on the fluid properties and the nominal diameter.
Page 106
Technical data Proline Promass E TB2 HART Weight [lb] [in] With transmitter housing GT20, With transmitter housing GT18 aluminum coating 1.4404/316L 13.2 18.7 ½ 14.3 19.8 17.6 23.2 1 ½ 28.7 34.2 48.5 54.0 Materials Transmitter housing • Compact version –...
Proline Promass E TB2 HART Technical data With transmitter housing GT18 1.4404/316L Electrical connection Type of protection Material Cable gland M20 × 1.5 • Non-Ex Stainless steel 1.4404 • Ex ia • Ex ic • Ex nA • Ex t Thread G ½"...
Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
• NAMUR NE 132 Coriolis mass meter Pressure Equipment Directive • With the PED/G1/III marking on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure Equipment Directive 97/23/EC. • Devices not bearing this marking (PED) are designed and manufactured according to good engineering practice.
Index Proline Promass E TB2 HART Index Symbols Connection tools ......25...
Page 111
Proline Promass E TB2 HART Index Commubox FXA191, 195 ....47 FieldCare ....... . 45 Commubox FXA291 .
Page 112
Index Proline Promass E TB2 HART Local display Mounting requirements see Measured value display Inlet and outlet runs ......19 see Diagnostics message Installation dimensions .
Page 113
Proline Promass E TB2 HART Index For the local display ..... . 59, 71 Sensor For totalizer ......70 Medium temperature range .
Page 114
Index Proline Promass E TB2 HART Supply voltage (Power supply) ....27, 99 Symbols Version data for the device ..... . 49 For channel number .
Page 115
Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.