5. Reconnect the device to the power supply.
 The configured device address is used once the device is restarted.
Setting the device name via the automation system
DIP switches 1-8 must all be set to OFF (factory setting) or all be set to ON to be able to set
the device name via the automation system.
The complete device name (name of station) can be changed individually via the
automation system.
• The serial number used as part of the device name in the factory setting is not
saved. It is not possible to reset the device name to the factory setting with the
serial number. The value "0" is used instead of the serial number.
• When assigning the device name via the automation system:
assign the device name in lower case letters.
7.6.2

Activating the default IP address

The default IP address 192.168.1.212 can be activated by DIP switch.
Activating the default IP address by DIP switch: Proline 500 - digital
Risk of electric shock when opening the transmitter housing.
Before opening the transmitter housing:
Disconnect the device from the power supply.
1.
1. Loosen the 4 fixing screws on the housing cover.
2. Open the housing cover.
3. Fold open the terminal cover.
4. Set DIP switch No. 2 on the I/O electronics module from OFF → ON.
5. Reverse the removal procedure to reassemble the transmitter.
6. Reconnect the device to the power supply.
 The default IP address is used once the device is restarted.