Write Protection Via Write Protection; Switch - Endress+Hauser Proline Promass E 500 PROFINET Operating Instructions Manual

Coriolis flowmeter
Table of Contents

Advertisement

Proline Promass E 500 PROFINET
Endress+Hauser
2. Define a max. 16-digit numeric code as an access code.
3. Enter the access code again in the Confirm access code parameter (→  153) to
confirm the code.
 The Web browser switches to the login page.
If no action is performed for 10 minutes, the Web browser automatically returns to
the login page.
• If parameter write protection is activated via an access code, it can also only be
deactivated via this access code →  80.
• The user role with which the user is currently logged on via Web browser is
indicated by the Access status parameter. Navigation path: Operation → Access
status
Resetting the access code
If you misplace the user-specific access code, it is possible to reset the code to the factory
setting. A reset code must be entered for this purpose. The user-specific access code can
then be defined again afterwards.
Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus
For a reset code, contact your Endress+Hauser service organization.
1. Navigate to the Reset access code parameter (→  154).
2. Enter the reset code.
 The access code has been reset to the factory setting 0000. It can be redefined
→  158.
10.8.2
Write protection via write protection switch
Unlike parameter write protection via a user-specific access code, this allows write access
to the entire operating menu - except for the "Contrast display" parameter - to be locked.
The parameter values are now read only and cannot be edited any more (exception
"Contrast display" parameter):
• Via local display
• Via PROFINET protocol
Proline 500 – digital
WARNING
L
Excessive tightening torque applied to the fixing screws!
Risk of damaging the plastic transmitter.
Tighten the fixing screws as per the tightening torque: 2 Nm (1.5 lbf ft)
Commissioning
159

Advertisement

Table of Contents
loading

Table of Contents