12.10 Event Logbook; Reading Out The Event Logbook - Endress+Hauser Proline Promass E 500 PROFINET Operating Instructions Manual

Coriolis flowmeter
Table of Contents

Advertisement

Proline Promass E 500 PROFINET
Endress+Hauser
 42
Taking the example of the local display
To call up the measures to rectify a diagnostic event:
• Via local display →  183
• Via Web browser →  184
• Via "FieldCare" operating tool →  185
• Via "DeviceCare" operating tool →  185

12.10 Event logbook

12.10.1 Reading out the event logbook

A chronological overview of the event messages that have occurred is provided in the
Events list submenu.
Navigation path
Diagnostics menu → Event logbook submenu → Event list
 43
Taking the example of the local display
• A maximum of 20 event messages can be displayed in chronological order.
• If the Extended HistoROM application package (order option) is enabled in the device,
the event list can contain up to 100 entries .
The event history includes entries for:
• Diagnostic events →  189
• Information events →  242
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
• : Occurrence of the event
• : End of the event
• Information event
: Occurrence of the event
To call up the measures to rectify a diagnostic event:
• Via local display →  183
• Via Web browser →  184
• Via "FieldCare" operating tool →  185
• Via "DeviceCare" operating tool →  185
For filtering the displayed event messages →  242
/ ../Diagnose list
Diagnostics
F273 Main electronic
Diagnostics 2
Diagnostics 3
/ ../Eventlist
I1091 Config. change
I1157 Mem.err. ev.list
0d01h19m10s
F311 Electr. failure
Diagnostics and troubleshooting
F
A0014006-EN
A0014008-EN