Summary of Contents for Endress+Hauser Proline Promass E 200
Page 1
Operating Instructions Proline Promass E 200 PROFIBUS PA Coriolis mass flow measuring system BA01133D/06/EN/01.12 71190789 Valid as of version 01.00.zz (Device firmware)
Page 2
Proline Promass E 200 PROFIBUS PA • Make sure the document is stored in a safe place such that it is always available when working on or with the device. • To avoid danger to individuals or the facility, read the "Basic safety instructions" section carefully, as well as all other safety instructions in the document that are specific to working procedures.
Proline Promass E 200 PROFIBUS PA Document information Document information Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
Document information Proline Promass E 200 PROFIBUS PA Allen key A0011221 Open-ended wrench A0011222 1.2.4 Symbols for certain types of information Symbol Meaning Allowed Indicates procedures, processes or actions that are allowed. A0011182 Preferred Indicates procedures, processes or actions that are preferred.
The following document types are available: • On the CD-ROM supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download For a detailed list of the individual documents along with the documentation code (®...
Observe the specified maximum process pressure. ► Verification for borderline cases: For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in ► verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
It meets general safety standards and legal requirements. It also complies with the EC directives listed in the device-specific EC Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. Endress+Hauser...
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany Ò TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA Ò Ò Ò Applicator , FieldCare , Field Xpert , HistoROM Registered or registration-pending trademarks of the Endress+Hauser Group Endress+Hauser...
Proline Promass E 200 PROFIBUS PA Incoming acceptance and product identification Incoming acceptance and product identification Incoming acceptance A0015502 A0013843 Is the order code on the delivery note (1) identical to the order code on the product sticker (2)? A0013695...
Incoming acceptance and product identification Proline Promass E 200 PROFIBUS PA If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. Product identification The following options are available for identification of the measuring device: • Nameplate specifications •...
Incoming acceptance and product identification Proline Promass E 200 PROFIBUS PA 4.2.3 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation.
Proline Promass E 200 PROFIBUS PA Storage and transport Storage and transport Storage conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections. They prevent mechanical damage to the sealing surfaces and fouling in the measuring tube.
Storage and transport Proline Promass E 200 PROFIBUS PA Packaging disposal All packaging materials are environmentally friendly and 100% recyclable: • Measuring device secondary packaging: polymer stretch film that conforms to EC Directive 2002/95/EC (RoHS). • Packaging: – Wood crate, treated in accordance with ISPM 15 standard, which is confirmed by the affixed IPPC logo.
Proline Promass E 200 PROFIBUS PA Installation Installation Installation conditions No special measures such as supports are necessary. External forces are absorbed by the construction of the device. 6.1.1 Mounting position Mounting location A0015595 To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: •...
Page 18
Installation Proline Promass E 200 PROFIBUS PA A0015596 Installation in a down pipe (e.g. for batching applications) å 4 Supply tank Sensor Orifice plate, pipe restriction Valve Batching tank Æ Orifice plate, pipe restriction [mm] [in] [mm] [in] 0.24 ½...
Proline Promass E 200 PROFIBUS PA Installation If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to the fluid properties. A0014057 Orientation of sensor with curved measuring tube å 5 Avoid this orientation for fluids with entrained solids: Risk of solids accumulating.
Page 20
Installation Proline Promass E 200 PROFIBUS PA For this reason, the following mounting locations are recommended: • At the lowest point in a vertical pipe • Downstream from pumps (no danger of vacuum) A0015594 Thermal insulation In the case of some fluids, it is important that the heat radiated from the sensor to the transmitter is kept to a minimum.
Proline Promass E 200 PROFIBUS PA Installation Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations. 6.1.3 Special mounting instructions Outlet run for peripheral device If a pressure and temperature measuring device are installed downstream from the measuring device, make sure there is sufficient distance between the two devices.
Installation Proline Promass E 200 PROFIBUS PA Mounting the measuring device 6.2.1 Required tools For transmitter • For turning the transmitter housing: Open-ended wrench8 mm • For opening the securing clamps: Allen key3 mm For sensor For flanges and other process connections: Corresponding mounting tools 6.2.2...
Proline Promass E 200 PROFIBUS PA Installation max. 350° 8 mm 8 mm A0013713 1. Release the fixing screw. 2. Turn the housing to the desired position. 3. Firmly tighten the securing screw. 6.2.5 Turning the display module 3 mm A0013905 1.
Page 24
Installation Proline Promass E 200 PROFIBUS PA Has the correct orientation for the sensor been selected (® ä 18)? • According to sensor type • According to fluid temperature • According to fluid properties (outgassing, with entrained solids) Does the arrow on the sensor nameplate match the direction of flow of the fluid through the piping (®...
Proline Promass E 200 PROFIBUS PA Electrical connection Electrical connection Connection conditions 7.1.1 Required tools • For cable entries: Use corresponding tools • For securing clamp: Allen key 3 mm • Wire stripper • When using stranded cables: Crimping tool for wire end ferrule •...
Page 26
Electrical connection Proline Promass E 200 PROFIBUS PA Cable type Cable structure Twisted, shielded twin-core cable Wire cross-section 0.8 mm (AWG 18) Loop resistance (direct 44 Ω/km current) Characteristic impedance 100 Ω ±20% at 31.25 kHz Attenuation constant at 3 dB/km 39.0 kHz...
Proline Promass E 200 PROFIBUS PA Electrical connection Bus termination The start and end of each fieldbus segment must always be terminated by a bus terminator. With various junction boxes (non-Ex), the bus termination can be activated via a switch. If this is not the case, a separate bus terminator must be installed.
Electrical connection Proline Promass E 200 PROFIBUS PA 7.1.5 Shielding and grounding Optimum electromagnetic compatibility (EMC) of the fieldbus system can only be guaranteed if the system components and, in particular, the lines are shielded and the shield forms as complete a cover as possible.
Proline Promass E 200 PROFIBUS PA Electrical connection NOTICE In systems without potential matching, the multiple grounding of the cable shield causes mains frequency equalizing currents! Damage to the bus cable shield. Only ground the bus cable shield to either the local ground or the protective ground at one ►...
Page 30
Electrical connection Proline Promass E 200 PROFIBUS PA Connection via terminals 3 mm 20 mm 10 (0.4) mm (in) A0013836 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Proline Promass E 200 PROFIBUS PA Electrical connection Removing a cable 3 (0.12) mm (in) A0013835 To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between ► the two terminal holes while simultaneously pulling the cable end out of the terminal.
Electrical connection Proline Promass E 200 PROFIBUS PA 2 + 8 = 10 A0015902 Example of hardware addressing; switch 8 is set to the "OFF" position; switches 1 to 7 define the address. å 7 Software addressing 1. Set switch 8 to "ON".
Page 33
Proline Promass E 200 PROFIBUS PA Electrical connection Do the cables have adequate strain relief? Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap" (® ä 32) ? Depending on the device version: are all the connectors firmly tightened (® ä 29)? Does the supply voltage match the specifications on the transmitter nameplate (®...
Operation options Proline Promass E 200 PROFIBUS PA Operation options Overview of operation options A0019091 Local operation via display module Computer with operating tool (e.g. FieldCare, SIMATIC PDM) Control system (e.g. PLC) Endress+Hauser...
Proline Promass E 200 PROFIBUS PA Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu with menus and parameters (® ä 134) Operating menu for operators and maintenances Language Display/operat.
Operation options Proline Promass E 200 PROFIBUS PA 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles. Each user role corresponds to typical tasks within the device lifecycle. Menu User role and tasks...
Proline Promass E 200 PROFIBUS PA Operation options Access to the operating menu via the local display 8.3.1 Operational display X X X X X X X 1120.50 kg/h A0016501 Operational display Device tag (® ä 65) Status area Display area for measured values (4-line) Operating elements (®...
Operation options Proline Promass E 200 PROFIBUS PA Mass flow A0013710 • Volume flow • Corrected volume flow A0013711 • Density • Reference density A0013946 Temperature A0013947 Totalizer The measurement channel number indicates which of the three totalizers is displayed.
Page 39
Proline Promass E 200 PROFIBUS PA Operation options Navigation path The navigation path - displayed at the top left in the navigation view - consists of the following elements: • In the submenu: Omission symbol for operating Name of current...
Operation options Proline Promass E 200 PROFIBUS PA Wizard A0013968 Parameters within a wizard No display symbol exists for parameters in submenus. A0013972 Locking Symbol Meaning Parameter locked When displayed in front of a parameter name, indicates that the parameter is locked.
Page 41
Proline Promass E 200 PROFIBUS PA Operation options Inserts minus sign at the input position. – A0016620 Confirms selection. A0013985 Moves the input position one position to the left. A0016621 Exits the input without applying the changes. A0013986 Clears all entered characters.
Operation options Proline Promass E 200 PROFIBUS PA Moves the input position one position to the left. A0013990 Deletes one character immediately to the left of the input position. A0013988 8.3.4 Operating elements Meaning Minus key In a menu, submenu Moves the selection bar upwards in a choose list.
Proline Promass E 200 PROFIBUS PA Operation options Meaning Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). A0013954 Minus/Plus/Enter key combination (press the keys simultaneously) For operational display Enables or disables the keypad lock.
Operation options Proline Promass E 200 PROFIBUS PA 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
Proline Promass E 200 PROFIBUS PA Operation options Navigation path "Expert" menu ® Direct access The direct access code consists of a 4-digit number and the channel number, which identifies the channel of a process variable: e.g. 0914-1. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
Operation options Proline Promass E 200 PROFIBUS PA 8.3.9 Changing the parameters For a description of the editing display - consisting of text editor and numeric editor - with symbols (® ä 40), for a description of the operating elements (® ä 42) Example: Changing the tag name in the "Tag description"...
Proline Promass E 200 PROFIBUS PA Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN 8.3.10 User roles and related access authorization The two user roles "Operator" and "Maintenance" have different write access to the parameters if the customer defines a user-specific access code.
Page 48
Operation options Proline Promass E 200 PROFIBUS PA XXXXXXXXXX 20.50 Keylock on kg/h A0014000-EN After disabling the keypad lock: XXXXXXXXXX 20.50 Keylock off kg/h A0014001-EN If the user attempts to access the operating menu while the keylock is enabled, the message Keylock on also appears.
Proline Promass E 200 PROFIBUS PA Operation options Access to the operating menu via the operating tool The structure of the operating menu in the operating tools is the same as for operation via the local display. 8.4.1 Connecting the operating tool...
FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
Proline Promass E 200 PROFIBUS PA Operation options User interface A0017402-EN Header Picture of device Device tag (® ä 65) Status area with status signal (® ä 92) Display area for current measured values (® ä 84) Event list with additional functions such as save/load, events list and document creation...
System integration Proline Promass E 200 PROFIBUS PA System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the Operating instructions • On transmitter nameplate (® ä 12) •...
Proline Promass E 200 PROFIBUS PA Commissioning Commissioning 10.1 Function check Before commissioning the device, make sure that the post-installation and post-connection checks have been performed. • "Post-mounting check" checklist (® ä 23) • "Post-connection check" checklist (® ä 32) 10.2...
Commissioning Proline Promass E 200 PROFIBUS PA X X X X X X X 20.50 0104-1 Main menu Language English Display/operat. Setup Main menu Display/operat. Setup Diagnostic / ../Setup Medium selection XXXXXXXXX XXXXXXXXX A0014007-EN Taking the example of the local display å...
Page 55
Proline Promass E 200 PROFIBUS PA Commissioning Volume flow unit Volume unit Corrected volume flow unit Corrected volume unit Density unit Reference density unit Temperature unit Length unit Pressure unit Parameter overview with brief description Parameter Description Selection/ Factory setting...
Page 56
Commissioning Proline Promass E 200 PROFIBUS PA Corrected volume unit Select the unit for standard Unit choose list Country-dependent: volume. • Nl • Scf Result The selected unit is taken from: Corrected volume flow unit Density unit Select the unit for density.
Proline Promass E 200 PROFIBUS PA Commissioning 10.4.2 Selecting and setting the medium The Medium selection wizard guides you systematically through all parameters that have to be configured for selecting and setting the medium. Navigation path "Setup" menu ® Medium selection...
Commissioning Proline Promass E 200 PROFIBUS PA Reference sound velocity Enter the sound velocity of 0 to 99 999 m/s 331.5 m/s the gas at 0°C (32°F). Temperature coefficient Enter the temperature Max. 15-digit, positive 0.61 (m/s)/K sound velocity coefficient of the sound floating-point number velocity of the gas.
Proline Promass E 200 PROFIBUS PA Commissioning 10.4.4 Configuring the analog inputs The Analog Inputs submenu guides the user systematically to the individual analog inputs 1-6. From here you get to the parameters of the individual analog input. Navigation path "Setup"...
Page 60
Commissioning Proline Promass E 200 PROFIBUS PA Structure of the wizard Format display Value 1 display 0% bargraph value 1 100% bargraph value 1 Value 2 display Value 3 display 0% bargraph value 3 100% bargraph value 3 Value 4 display...
Page 61
Proline Promass E 200 PROFIBUS PA Commissioning 0% bargraph value 1 Enter the 0% value to be Floating point number with Country-dependent: shown on the bargraph sign • 0 kg/h display for the measured • 0 lb/min value 1. 100% bargraph value 1...
Commissioning Proline Promass E 200 PROFIBUS PA 10.4.6 Configuring the low flow cut off The Low flow cut off wizard guides you systematically through all parameters that have to be set for configuring the low flow cut off. Navigation path "Setup"...
Proline Promass E 200 PROFIBUS PA Commissioning 10.4.7 Configuring the partial filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation path "Setup"...
Commissioning Proline Promass E 200 PROFIBUS PA 10.5 Advanced settings The Advanced setup menu with its submenus contains all parameters needed for specific settings. Navigation path "Setup" menu ® Advanced setup Navigation to the "Advanced setup" submenu X X X X X X X 20.50...
Proline Promass E 200 PROFIBUS PA Commissioning Totalizer 1 ® (® ä 76) ® Totalizer 2 (® ä 76) ® Totalizer 3 (® ä 76) ® Configuration backup (® ä 77) display 10.5.1 Defining the tag name To enable quick identification of the measuring point within the system, you can enter a unique designation using the Device tag parameter and thus change the factory setting.
Proline Promass E 200 PROFIBUS PA Commissioning Installation direction ® Zero point adjustment Zero point adjustment control Adjustment in progress Parameter overview with brief description Parameters Description Selection/ Factory setting User entry Installation direction Change the sign of the • Flow in arrow direction Flow in arrow direction direction of flow of the fluid.
Commissioning Proline Promass E 200 PROFIBUS PA Structure of the wizard for the pulse output Operating mode Pulse Assign pulse Volume flow Correct.vol.flow Mass flow Mass unit Volume unit Corr. vol. unit Corr. vol.flow. Calc ref density Fix ref.density Linear exp coeff Ref.
Page 69
Proline Promass E 200 PROFIBUS PA Commissioning Assign pulse output Select the process variable • Off Mass flow for the pulse output. • Mass flow • Volume flow • Corrected volume flow Mass unit Select the unit for mass. Unit choose list Country-dependent: •...
Page 70
Commissioning Proline Promass E 200 PROFIBUS PA Reference temperature Enter the reference Max. 15-digit floating-point Country-dependent: • 20 °C temperature for calculating number with sign the reference density. • 68 °F Pulse value Enter the measured value Depends on the process for the pulse output.
Page 71
Proline Promass E 200 PROFIBUS PA Commissioning Structure of the wizard for the frequency output Operating mode Frequency Assign freq. Volume flow Density Temperature Ref.density Mass flow Correct.vol.flow Temperature Volume flow Density unit unit unit Mass flow Corr. vol. unit Ref.
Page 72
Commissioning Proline Promass E 200 PROFIBUS PA Assign frequency output Select the process variable • Off for the frequency output. • Mass flow • Volume flow • Corrected volume flow • Density • Reference density • Temperature Mass flow unit...
Page 73
Proline Promass E 200 PROFIBUS PA Commissioning Temperature unit Select temperature unit. Unit choose list Country-dependent: • °C (Celsius) Result • °F (Fahrenheit) The selected unit applies for: – Outputs – Reference temperature – Simulation process variable Minimum frequency value Enter the minimum 0.0 to 1 000.0...
Page 74
Commissioning Proline Promass E 200 PROFIBUS PA Operating mode Switch Switch out funct Fl. direct. Limit Status check Assign Assign Assign limit dir.check status Ref. Partial Mass Correct. Totalizer Mass Volume Correct. Low flow density pipe det. flow vol.flow 1/2/3...
Page 75
Proline Promass E 200 PROFIBUS PA Commissioning Switch output function Select the function for the • Off switch output. • On • Diagnostic behavior • Limit value • Flow direction monitoring • Status Assign diagnostic behavior Select the diagnostic • Alarm...
Commissioning Proline Promass E 200 PROFIBUS PA Density unit Select the unit for density. Unit choose list Country-dependent • kg/l Result • lb/cf The selected unit applies for: – Outputs – Low value partial filled pipe detection – High value partial filled pipe detection –...
Proline Promass E 200 PROFIBUS PA Commissioning Navigation path "Setup" menu ® Advanced setup ® Totalizer 1-3 Structure of the submenu ® Totalizer 1-3 Assign process variable Unit totalizer Control totalizer Operating mode Failure mode Parameter overview with brief description...
Page 78
Commissioning Proline Promass E 200 PROFIBUS PA Navigation path "Setup" menu ® Advanced setup ® Conf. backup disp. Function scope of the "Configuration management" parameter Options Description Execute backup The current device configuration is backed up from the integrated HistoROM to the device's display module.
Proline Promass E 200 PROFIBUS PA Commissioning Comparison result Comparison between • Set. identical Check not done present device data and • Set. not ident. display backup • No backup • Backup corrupt • Check not done • Dataset incomp.
Commissioning Proline Promass E 200 PROFIBUS PA Value process One of the following options is Enter the Depends on the variable selected in the Assign simulation value for process variable simulation process variable the selected process selected parameter: variable. • Mass flow •...
Proline Promass E 200 PROFIBUS PA Commissioning Navigation path "Setup" menu ® Advanced setup ® Define access code Structure of the submenu ® Define access code Define access code Confirm access code Defining the access code via local display 1. Navigate to the Define access code parameter.
Page 82
Commissioning Proline Promass E 200 PROFIBUS PA The parameter values are now read only and cannot be edited any more (exception Contrast display): • Via local display • Via PROFIBUS PA protocol XXXXXXXXXXXXX 3 mm A0013768 1. Loosen the securing clamp.
Page 83
Proline Promass E 200 PROFIBUS PA Commissioning X X X X X X X 20.50 A0015870 If hardware write protection is disabled, no option is displayed in the Locking status parameter (® ä 84). On the local display, the -symbol disappears from in front of the parameters in the header of the operational display and in the navigation view.
Operation Proline Promass E 200 PROFIBUS PA Operation 11.1 Read device locking status The write protection types that are currently active can be determined using the Locking status parameter. Navigation path "Expert" menu ® Locking status Function scope of "Locking status" parameter...
Proline Promass E 200 PROFIBUS PA Operation Density Reference density Temperature Parameter overview with brief description Parameter overview with brief description Parameter Prerequisite Description Display Mass flow – Displays the mass flow Floating point number currently measured with sign Volume flow –...
Operation Proline Promass E 200 PROFIBUS PA Totalizer value – Displays the current Floating point number totalizer counter value. with sign Totalizer status – Displays the totalizer • GOOD status. • BAD • UNCERTAIN Status (hex) – Information on the...
Proline Promass E 200 PROFIBUS PA Operation 11.6 Performing a totalizer reset Navigation path "Display/operat." menu ® Operation Function scope of the "Control totalizer" parameter Options Description Totalize The totalizer is started. Reset + Hold The totaling process is stopped and the totalizer is reset to 0.
Page 88
Operation Proline Promass E 200 PROFIBUS PA Function scope • A total of 1000 measured values can be stored • 4 logging channels • Adjustable logging interval for data logging • Display of the measured value trend for each logging channel in the form of a chart...
Page 89
Proline Promass E 200 PROFIBUS PA Operation Assign channel 1-4 Assign process • Off variable to logging • Mass flow channel. • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency •...
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting 12.1 General troubleshooting For local display Problem Possible cause Remedy Local display dark and no output Supply voltage does not match that Apply the correct supply voltage signals specified on the nameplate.
Page 91
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting No write access to parameters Current user role has limited access 1. Check user role (® ä 47). authorization 2. Enter correct customer-specific access code (® ä 47). No connection via service interface...
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA 12.2 Diagnostic information on local display 12.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
Page 93
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. A0013961 • For local display with touch control: the background lighting changes to red.
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA 12.2.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage...
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting 12.3 Diagnostic information in FieldCare 12.3.1 Diagnostic options Any faults detected by the measuring device are displayed on the home page of the operating tool once the connection has been established.
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Navigation path "Expert" menu ® System ® Diagnostic handling ® Condensed status ®Assign behavior of measured value status xx 1421-1 / ../Condensed statu. Var. status: 44 Uncert.:Proc.(S) Dev.status: 44 Var. status: 46...
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting Navigation path "Expert" menu ® System ® Diagnostic handling ® Diagnostic behavior ® Assign behavior of diagnostic no. xxx 0658-1 / ../Diagn. behavior Diagnostic no.442 Warning Diagnostic no.443 A0019179-EN You can assign the following options to the diagnostic number as the diagnostic behavior:...
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Diagnostic no. 046 PROFIBUS measured value status (Profile 3.02) Measured variables concerned Message: *046 Sensor limit • Density • Mass flow Substatus Process related, no maintenance • Reference density • Corrected volume flow...
Page 99
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting • Temperature Quality: • Volume flow Local display Remedial measures Message: F270 Main electronic failure Change main electronics. Diagnostic behavior: Alarm Diagnostic no. 271 PROFIBUS measured value status (Profile 3.02) Measured variables concerned...
Page 100
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Status (hex): 0x78 Status signal: Quality: UNCERTAIN Local display Remedial measures Message: S274 Main electronic failure Unstable measurement: Change main electronics. Diagnostic behavior Warning Configurable substatus; Configurable diagnostic behavior(® ä 95) Diagnostic no.
Page 101
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting Diagnostic no. 311 PROFIBUS measured value status (Profile 3.02) Measured variables concerned Message: F311 Electronic failure • Density • Partially filled pipe detection Substatus: Maintenance alarm, more diagnosis available • Low flow cut off •...
Page 102
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Local display Remedial measures Message: C411 Up-/download active Up-/download active: Wait until the up-/download is complete. Diagnostic behavior: Warning Diagnostic no. 438 PROFIBUS measured value status (Profile 3.02) Measured variables concerned...
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting • Low flow cut off Substatus: Function check • Mass flow Status (hex): 0xBC • Pulse frequency switch output status • Reference density Status signal: • Corrected volume flow • Temperature...
Page 104
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Message: S831 Ambient temperature Reduce ambient temperature in the area of the sensor housing. Diagnostic behavior Warning Configurable substatus; Configurable diagnostic behavior(® ä 95) Diagnostic no. 832 PROFIBUS measured value status (Profile 3.02)
Page 105
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting • Reference density Substatus Process related, no maintenance • Corrected volume flow Status (hex): 0x78 • Temperature Status signal: Quality: UNCERTAIN Local display Remedial measures Message: S835 Process temperature Increase process temperature.
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA • Increase system pressure. Diagnostic behavior Warning In particular with outgassing media and/or increased gas content, the following measures are recommended to increase system pressure: • Mount the device at the outlet side of a pump.
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting Other diagnostic events that are pending can be viewed in the Diagnostic list submenu (® ä 107) 12.7 Diagnostic list In the Diagnostic list submenu, up to 5 currently pending diagnostic events can be displayed along with the related diagnostic information.
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA • Out of specification (S) • Maintenance required (M) • Information (I) 12.8.3 Overview of information events Unlike a diagnostic event, an information event is displayed in the event logbook only and not in the diagnostic list.
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting I1397 Fieldbus: access status modified I1398 CDI: access status modified 12.9 Resetting the measuring device Using the Device reset parameter it is possible to reset the entire device configuration or some of the configuration to a defined state.
Diagnostics and troubleshooting Proline Promass E 200 PROFIBUS PA Serial number Displays the serial number Max. 11-digit character of the measuring device. string comprising letters and numbers The number can be found on the nameplate of the sensor and transmitter.
Page 111
Proline Promass E 200 PROFIBUS PA Diagnostics and troubleshooting The manufacturer's information is available: • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download • Specify the following details: – Product root, e.g. 8E2B – Text search: Manufacturer's information –...
• Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices by Endress+Hauser Service or at the factory only.
• Is located on the device nameplate and the spare part overview sign. • Can be read out via the Serial number parameter in the Device information submenu (® ä 109). 13.3 Endress+Hauser services Contact your Endress+Hauser Sales Center for information on services and spare parts. Endress+Hauser...
• Observe the maximum permitted fluid temperature for the measuring device (® ä 126). 14.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services.
Proline Promass E 200 PROFIBUS PA Return Return The measuring device must be returned if repairs or a factory calibration are required, or if the wrong measuring device has been ordered or delivered. According to legal regulations, Endress +Hauser, as an ISO-certified company, is required to follow certain procedures when handling returned products that are in contact with medium.
Disposal Proline Promass E 200 PROFIBUS PA Disposal 16.1 Removing the measuring device 1. Switch off the device. 2. WARNING! Danger to persons from process conditions. Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
Proline Promass E 200 PROFIBUS PA Technical data Technical data 17.1 Application The measuring device is suitable for flow measurement of liquids and gases only. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
Technical data Proline Promass E 200 PROFIBUS PA Maximum full scale value for gas [kg/h] max(G) Maximum full scale value for liquid [kg/h] max(F) < g can never be greater than g max(G) max(F) max(G) max(F) Gas density in [kg/m...
Page 119
Proline Promass E 200 PROFIBUS PA Technical data Maximum input values • DC 35 V • 50 mA For information on the Ex connection values • For £ 2 mA: 2 V Voltage drop • For 10 mA: 8 V £...
Page 120
Technical data Proline Promass E 200 PROFIBUS PA Pulse/frequency/switch output Pulse output Failure mode Choose from: • Actual value • No pulses Frequency output Failure mode Choose from: • Actual value • Defined value: 0 to 1 250 Hz • 0 Hz...
Proline Promass E 200 PROFIBUS PA Technical data Output values Analog input 1 to 6 (from measuring device to • Mass flow automation system) • Volume flow • Corrected volume flow • Density • Reference density • Temperature Digital input 1 to 2 •...
Page 122
Technical data Proline Promass E 200 PROFIBUS PA Power consumption Transmitter Order code for Maximum power consumption "Output" Option G: PROFIBUS PA, pulse/frequency/switch output • Operation with output 1: 512 mW • Operation with output 1 and 2: 2 512 mW...
Proline Promass E 200 PROFIBUS PA Technical data 17.6 Performance characteristics Reference operating • Error limits based on ISO 11631 • Water with +15 to +45 °C (+59 to +113 °F) at 2 to 6 bar (29 to 87 psi) conditions •...
Page 124
Technical data Proline Promass E 200 PROFIBUS PA Design fundamentals (® ä 125) Flow values Flow values as turndown parameter depending on nominal diameter. SI units 1:10 1:20 1:50 1:100 1:500 [mm] [kg/h] [kg/h] [kg/h] [kg/h] [kg/h] [kg/h] 2 000...
Proline Promass E 200 PROFIBUS PA Technical data Influence of medium Mass flow and volume flow temperature When there is a difference between the temperature for zero point adjustment and the process temperature, the typical measured error of the sensor is ±0.0002 % of the full scale value/°C (±0.0001 % of the full scale value/°F).
Technical data Proline Promass E 200 PROFIBUS PA –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) Storage temperature Climate class DIN EN 60068-2-38 (test Z/AD) Transmitter Degree of protection • As standard: IP66/67, type 4X enclosure •...
Page 127
Proline Promass E 200 PROFIBUS PA Technical data Flow limit Select the nominal diameter by optimizing between the required flow range and permissible pressure loss. For an overview of the measuring range full scale values, see the "Measuring range" section (®...
Technical data Proline Promass E 200 PROFIBUS PA [mbar] 10000 DN 40 DN 25 DN 15 DN 50 DN 8 1000 0.01 0.001 1000 [t/h] A0013674 Pressure loss diagram for water å 29 Pressure loss in US units Pressure loss depends on the fluid properties and the nominal diameter.
Page 129
Proline Promass E 200 PROFIBUS PA Technical data Weight [lbs] [in] Order characteristic for "Housing", Order characteristic for "Housing", option C option B Aluminum coating 1.4404/316L 17.6 23.2 1 ½ 28.7 34.2 48.5 54.0 Materials Transmitter housing • Order code for "Housing", option B: stainless steel 1.4404/316L •...
Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
Page 132
• The device can also be operated with certified devices of other manufacturers (interoperability) Pressure Equipment Directive • With the identification PED/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms conformity with the "Basic Safety Requirements" specified of Appendix I of the Pressure Equipment Directive 97/23/EC.
For an overview of the accessories that can be ordered, see the "Technical Information" document 17.15 Documentation The following document types are available: • On the CD-ROM supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download Standard documentation Device type: 8E2B**- Communication...
Appendix Proline Promass E 200 PROFIBUS PA Appendix 18.1 Overview of the Operator/Maintenance operating menu The following table provides an overview of the entire operating menu structure with menus and parameters. The direct access code to the parameter is given in brackets. The page reference indicates where a description of the parameter can be found in the manual.
Page 135
Proline Promass E 200 PROFIBUS PA Appendix Pressure unit (® ä 56) ® Select medium (® ä 57) Select medium (® ä 57) Select gas type (® ä 57) Reference sound velocity (® ä 58) Temperature coefficient (® ä 58)
Page 136
Appendix Proline Promass E 200 PROFIBUS PA Pressure shock suppression (® ä 62) ® Partial filled pipe (® ä 63) detection Assign process variable (® ä 63) Low value (® ä 63) High value (® ä 63) Response time part. filled (®...
Page 137
Proline Promass E 200 PROFIBUS PA Appendix Volume flow unit Volume unit Corrected volume flow unit Corrected volume unit Density unit Unit totalizer 1 to 3 Reference density unit Temperature unit Value per pulse (0455) (® ä 70) Pulse width (0452) (®...
Page 138
Appendix Proline Promass E 200 PROFIBUS PA Failure mode (® ä 77) ® Conf. backup disp. (® ä 77) Operating time (0652) (® ä 78) Last backup (0102) (® ä 78) Configuration management (® ä 78) (0100) Comparison result (0103) (®...
Page 139
Proline Promass E 200 PROFIBUS PA Appendix Reference density (® ä 85) Temperature (® ä 85) ® Totalizer 1 to 3 (® ä 85) Assign process variable (® ä 86) Totalizer value (® ä 86) Totalizer status Status (hex) ®...
Page 141
Proline Promass E 200 PROFIBUS PA Appendix Display interval (0096) Display damping (0094) Header (0097) Header text (0112) Separator (0101) Contrast display (0105) (® ä 42) Backlight (0111) Access status display (® ä 47) (0091) Conf. backup disp. ® (® ä 77) Operating time (0652) (®...
Page 142
Appendix Proline Promass E 200 PROFIBUS PA Variable status for diagnostic no. 831 (1419) Device status for diagnostic no. 831 (1403) Variable status for diagnostic no. 832 (1422) Device status for diagnostic no. 832 (1406) Variable status for diagnostic no. 833 (1423) Device status for diagnostic no.
Page 144
Appendix Proline Promass E 200 PROFIBUS PA Temperature unit (0557) (® ä 56) Length unit (0551) (® ä 56) Pressure unit (0564) (® ä 56) Date/time format (2812) ® User spec. units User mass text (0560) User mass offset (0562)
Page 145
Proline Promass E 200 PROFIBUS PA Appendix Response time (1836) (® ä 63) Max. damping (2492) ® Measuring mode Select medium (2513) (® ä 57) Select gas type (2488) (® ä 57) Reference sound velocity (® ä 58) (2489) Temperature coefficient (®...
Page 146
Appendix Proline Promass E 200 PROFIBUS PA Volume flow offset (1815) Volume flow factor (1821) Corrected volume flow offset (1824) Corrected volume flow factor (1823) Density offset (1826) Density factor (1825) Reference density offset (1828) Reference density factor (1827) Temperature offset (1830) Temperature factor (1829) ®...
Page 147
Proline Promass E 200 PROFIBUS PA Appendix Assign frequency output (® ä 72) (0478) Minimum frequency value (® ä 73) (0453) Maximum frequency value (® ä 73) (0454) Measuring value at (® ä 73) minimum frequency (0476) Measuring value at (®...
Page 148
Appendix Proline Promass E 200 PROFIBUS PA Ident number selector (1461) ® PROFIBUS PA info Status PROFIBUS Master Config (1465) PROFIBUS ident number (1464) Profile version (1463) Base current (1466) Terminal voltage 1 (0662) ® Physical block Tag description (1496)
Page 149
Proline Promass E 200 PROFIBUS PA Appendix Device install date (1491) Ident number selector (1461) Hardware lock (1499) Feature supported (1477) Feature enabled (1476) Condensed status diagnostic (1500) ® Analog inputs Analog input 1 to 6 ® Tag description (1562)
Page 150
Appendix Proline Promass E 200 PROFIBUS PA Channel (1561) Out unit (1550) Out decimal point (1547) PV filter time (1524) Fail safe type (1525) Fail safe value (1526) Alarm hysteresis (1527) Hi Hi Lim (1528) Hi Lim (1529) Lo Lim (1530)
Page 151
Proline Promass E 200 PROFIBUS PA Appendix Mode block permitted (1648) Mode block normal (1643) Alarm summary (1642) Batch-ID (1633) Batch operation (1639) Batch phase (1640) Batch Recipe Unit Procedure (1641) Set point value (1661) Set point status (1660) PV scale lower range...
Page 152
Appendix Proline Promass E 200 PROFIBUS PA Out status (1669) Out status (1645) Out scale upper range (1646) Out scale lower range (1644) ® Digital inputs ® Digital input 1 to 2 Tag description (2201) Static revision (2200) Strategy (2199)
Page 153
Proline Promass E 200 PROFIBUS PA Appendix Digital output 1 to 3 Tag description (1721) Static revision (1720) Strategy (1719) Alert key (1694) Target mode (1722) Mode block actual (1691) Mode block permitted (1705) Mode block normal (1702) Alarm summary (1701)
Page 154
Appendix Proline Promass E 200 PROFIBUS PA Simulate enabled (1716) Simulate value (1718) Simulate status (1717) Application Totalizer 1 to 3 ® (® ä 76) Tag description (3833) Static revision (3832) Strategy (3831) Alert key (3803) Target mode (3834) Mode block actual (3801)
Page 155
Proline Promass E 200 PROFIBUS PA Appendix Hi Lim (3816) Lo Lim (3819) Lo Lo Lim (3822) Hi Hi Alarm value (3814) Hi Hi Alarm state (3813) Hi Alarm value (3812) Hi Alarm state (3811) Lo Alarm value (3818) Lo Alarm state (3817)
Page 156
Appendix Proline Promass E 200 PROFIBUS PA Data logging ® (® ä 87) Assign channel 1-4 (® ä 89) (851 to 854) Logging interval (0856) (® ä 89) Clear logging data (0855) (® ä 89) Display channel 1-4 (® ä 89) ®...
Page 157
Proline Promass E 200 PROFIBUS PA Appendix ® Oscillation damping Minimum value (2586) Maximum value (2585) Signal asymmetry ® Minimum value (2551) Maximum value (2476) ® Simulation (® ä 79) Assign simulation process (® ä 79) variable (1810) Value process variable (®...