Installation
28
WARNING
L
Danger from medium escaping!
Medium escaping under pressure can cause injury or material damage.
Take precautions to prevent danger to persons and damage if the rupture disk is
actuated.
Observe information on the rupture disk sticker.
Make sure that the function and operation of the rupture disk is not impeded through
the installation of the device.
Do not use a heating jacket.
Do not remove or damage the rupture disk.
After the rupture disk is actuated, do not operate the measuring device any more.
The position of the rupture disk is indicated on a sticker applied over it. If the rupture disk
is triggered, the sticker is destroyed. The disk can therefore be visually monitored.
DN 8 ( ")...
1
Rupture disk label
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology.
Calibration takes place under reference conditions→  264. Therefore, a zero point
adjustment in the field is generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates.
• Under extreme process or operating conditions (e.g. very high process temperatures or
very high-viscosity fluids).
Protective cover
213 (8.4)
 10
Weather protection cover for Proline 500 – digital
50 2
( ")
RUPTURE DISK
i
1
Proline Promass E 500 PROFINET
DN
80 3
( ")
203 (8.0)
Endress+Hauser
A0029956
A0029552