Testing The Module; Performance Test Record - Agilent Technologies 16962A Service Manual

Logic analyzer
Table of Contents

Advertisement

3
Testing Logic Analyzer Performance

Testing the Module

N O T E
40
The following sections explain how to test the maximum
state data rate.
Record the 16962A logic analyzer's model and serial
1
number in the Performance Test Record (see
Record your work order number (if applicable) and
today's date.
Record the test equipment information in the "Test
2
Equipment Used" section of the Performance Test Record.
Turn on the logic analysis system.
3
Before testing the performance of the module, warm-up the logic analyzer
and the test equipment for 30 minutes.
Optionally, connect a keyboard and mouse to the rear
a
panel of the mainframe.
b
Plug in the power cord to the power connector on the
rear panel of the mainframe.
Turn on the main power switch on the mainframe
c
front panel.
While the logic analysis system is booting, observe the
boot dialogue for the following:
Ensure all of the installed memory is recognized.
Any error messages.
Interrupt of the boot process with or without error
message.
4
During initialization, check for any failures.
If an error or an interrupt occurs, refer to the Agilent
Technologies 16900- Series Logic Analysis System Service
Guide for troubleshooting information.
page
16962A Logic Analyzer Service Guide
59).