Replacing The Rear Panel Power Reference Cable Assembly (Option 003) - Agilent Technologies E4418B Service Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

Replacing the Rear Panel Power Reference Cable Assembly
(Option 003)
1.
Disassemble the power meter down to allow access to the rear panel
connectors. (Refer to "Removing the A4 Comms Assembly", on page 5-11.)
2.
Replace the power reference cable assembly using the replacement kit and
the torques indicated on the following diagrams.
6 lb/in
Agilent E4418B/E4419B Service Guide
Assembly and Disassembly Guidelines
tie wraps
Replaceable Parts
25 lb/in
9 lb/in
5-21