Testing the Module Using a 16900-Series Mainframe

The following sections explain how to test the minimum master-to-master clock
time and minimum eye width.
Use the following instructions to test the module using a 16900-series mainframe.
If you are using a 16700-series mainframe, use the instructions beginning on page
61.
1 Record the 16753A, 16754A, 16755A, 16756A, or 16950A logic analyzer's
model and serial number in the Performance Test Record (see page 60).
Record your work order number (if applicable) and today's date.
2 Record the test equipment information in the "Test Equipment Used"
section of the Performance Test Record.
3 Turn on the logic analysis system.
Before testing the performance of the module, warm-up the logic analyzer
NOTE:
and the test equipment for 30 minutes.
a Connect the keyboard and monitor to the rear panel of the logic
b Connect the mouse to the rear panel of the mainframe.
c Plug in the power cord to the power connector on the rear panel of the
d Turn on the main power switch on the mainframe front panel.
4 During initialization, check for any failures.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
analysis mainframe (16900A only).
mainframe.
While the logic analysis system is booting, observe the boot dialogue for the following:
• ensure all of the installed memory is recognized
• any error messages
• interrupt of the boot process with or without error message
If an error or an interrupt occurs, refer to the Agilent Technologies 16900A-series
Logic Analysis System Service Guide for troubleshooting information.
Chapter 3: Testing Logic Analyzer Performance
Testing the Module Using a 16900-Series Mainframe
37