Analog Motherboard Replacement; Tools Required; Removal Procedure - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Replacement Procedure

Analog Motherboard Replacement

Analog Motherboard Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)

Removal Procedure

Refer to Figure 6-27 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 266.
Step 2. Remove the following assemblies.
Front Panel Assembly (refer to "Front Panel Removal" on page 267.)
Removable Hard Disk Assembly (refer to "Removable Hard Disk Assembly Removal"
on page 268.)
Level Vernier Module (refer to "Level Vernier Module Replacement" on page 296.)
Synthesizer Module (refer to "Synthesizer Module Replacement" on page 298.)
Receiver Module (refer to "Receiver Module Replacement for 2-Port model" on
page 300 or "Receiver Module Replacement for 4-Port model" on page 302.)
Step 3. Remove the two TORX T10 screws (item 3) fastening the plate and remove it.
Step 4. Disconnect the flat cables connected to the analog motherboard (item 1, 2).
Step 5. Remove the six TORX T10 screws (item 4) fastening the rear cover.
Step 6. Disconnect the cable (item 5) connected to the analog motherboard when with option 1E5.
Step 7. Disconnect the cables (item 6, 7) connected to the analog motherboard.
Step 8. Remove the ten TORX T10 screws (item 8) fastening the analog mother board.
312
Chapter 6

Advertisement

Table of Contents
loading

Table of Contents