Agilent Technologies E5071C Service Manual page 55

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

With Option 1E5
Frequency [Hz]
Test limit [Hz]
9 k
100 k
50 M
134.1 M
± 0.548 k
548 M
3 G
8.5 G
Option 285, 485
Without Option 1E5
Frequency [Hz]
Test limit [Hz]
100 k
50 M
134.1 M
548 M
3 G
8.5 G
With Option 1E5
Frequency [Hz]
Test limit [Hz]
100 k
50 M
134.1 M
± 0.548 k
548 M
3 G
8.5 G
Chapter 2
E5071C Performance Test Record
Test result [Hz]
± 0.009
± 0.1
± 50
± 134.1
± 3 k
± 8.5 k
Test result [Hz]
± 0.5
± 250
± 670.5
± 2.74 k
± 15 k
± 42.5 k
Test result [Hz]
± 0.1
± 50
± 134.1
± 3 k
± 8.5 k
Performance Test
Measurement
uncertainty [Hz]
± 0.0003
± 0.003
± 0.58
± 1.6
± 6.4
± 35
± 99
Measurement
uncertainty [Hz]
± 0.004
± 1.7
± 4.6
± 16.4
± 90
± 254
Measurement
uncertainty [Hz]
± 0.003
± 0.58
± 1.6
± 6.4
± 35
± 99
55

Advertisement

Table of Contents
loading

Table of Contents