Agilent Technologies E5071C Service Manual page 204

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Troubleshooting
Performance test failure troubleshooting
Table 4-7
Performance test failure troubleshooting information
First failed test
System Dynamic Range
(S32)
System Dynamic Range
(S43)
System Dynamic Range
(S14, S12)
Dynamic Accuracy (T1)
Dynamic Accuracy (T2)
Dynamic Accuracy (T3)
Dynamic Accuracy (T4)
Uncorrected System
Performance (S1, D1, R1)
Uncorrected System
Performance (S2, D2, R2)
Uncorrected System
Performance (S3, D3, R3)
Uncorrected System
Performance (S4, D4, R4)
Uncorrected System
Performance
(L21,L12,T21,T12)
Uncorrected System
Performance
(L31,L13,T31,T13)
Uncorrected System
Performance
(L41,L14,T41,T14)
Uncorrected System
Performance
(L32,L23,T32,T23)
Uncorrected System
Performance
(L42,L24,T42,T24)
Uncorrected System
Performance
(L43,L34,T43,T34)
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
204
Probable faulty board assembly or parts
OCXO
BAS
SYN
SYN
-SRC
-LCL
##
##
##
##
##
##
#
#
#
#
#
#
#
#
##
##
##
##
##
##
##
##
#
#
#
#
#
#
#
#
#
#
#
#
VNR
RCV
RCV
RCV
-1
-2
-3
#
###
#
#
###
#
###
#
###
#
###
#
#
###
#
###
#
###
#
###
###
###
###
###
###
###
###
###
RCV
T2
CBL
-4
-SW
#
###
#
#
#
#
#
###
#
#
#
#
###
#
##
##
###
##
##
###
##
###
##
Chapter 4