Performance Test
E5071C Performance Test Calculation Sheet
Name
A1
A2
Power
setting
[dB]
DA45
5
40
( Tx @ DA45) - ( Tx @ DA40b) - (Pm @ DA45) + (Pm @ DA40b) →
True D.A. = (Running D.A. [dB] @ DA45) + (True D.A. [dB] @ DA40)
DA50
10
40
( Tx @ DA50) - ( Tx @ DA40b) - (Pm @ DA50) + (Pm @ DA40b) →
True D.A. = (Running D.A. [dB] @ DA50) + (True D.A. [dB] @ DA40)
DA50a
0
50
Adjust source power for (Pm [db] @ DA50a) - (Tx [dB] @ DA50a) + ( Tx [dB] @ DA50)
DA50b
0
50
DA55
5
50
( Tx @ DA55) - ( Tx @ DA50b) - (Pm @ DA55) + (Pm @ DA50b) →
True D.A. = (Running D.A. [dB] @ DA55) + (True D.A. [dB] @ DA50)
DA60
10
50
( Tx @ DA60) - (Tx @ DA50b) - (Pm @ DA60) + (Pm @ DA50b) →
True D.A. = (Running D.A. [dB] @ DA60) + (True D.A. [dB] @ DA50)
DA60a
0
60
Adjust source power for (Pm [db] @ DA60a) - ( Tx [dB] @ DA60a) + ( Tx [dB] @ DA60)
DA60b
0
60
DA65
5
60
( Tx @ DA65) - ( Tx @ DA60b) - (Pm @ DA65) + (Pm @ DA60b) →
True D.A. = (Running D.A. [dB] @ DA65) + (True D.A. [dB] @ DA60)
DA70
10
60
( Tx @ DA70) - ( Tx @ DA60b) - (Pm @ DA70) + (Pm @ DA60b) →
True D.A. = (Running D.A. [dB] @ DA70) + (True D.A. [dB] @ DA60)
DA70a
0
70
Adjust source power for (Pm [db] @ DA70a) - ( Tx [dB] @ DA70a) + ( Tx [dB] @ DA70)
DA70b
0
70
DA75
5
70
( Tx @ DA75) - ( Tx @ DA70b) - (Pm @ DA75) + (Pm @ DA70b) →
True D.A. = (Running D.A. [dB] @ DA75) + (True D.A. [dB] @ DA70)
DA80
10
70
( Tx @ DA80) - ( Tx @ DA70b) - (Pm @ DA80) + (Pm @ DA70b) →
True D.A. = (Running D.A. [dB] @ DA80) + (True D.A. [dB] @ DA70)
DA80a
0
80
48
Pm [dB]
Tx [dB] @ Port (Port Char
OFF)
Running
True
D.A.
D.A.
———
———
———
———
———
———
REF50
REF50
———
———
———
———
———
———
REF60
REF60
———
———
———
———
———
———
REF70
REF70
———
———
———
———
———
———
Chapter 2