PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B

Performance Tests (Agilent N7840A Software Package)

The Agilent N7840A software package verifies the electrical performance of your E8364B
PNA series microwave network analyzer. The software automatically configures your
analyzer to execute the performance tests. The N7840A software package is not included
with the analyzer; it must be ordered separately. The model numbers of the equipment
used are specified under
There are nine tests in the software package:
Source Power Linearity Test
Source Maximum Power Output Test

Source Power Accuracy Test

Frequency Accuracy Test
Trace Noise Test
Receiver Compression Test
Noise Floor Test
Calibration Coefficient Test
System Crosstalk Test
Dynamic Accuracy Test
IMPORTANT
The descriptions of these tests are provided here as a convenience only. These
tests are subject to change at any time. As a result of such change, the
descriptions provided here may no longer be accurate. Refer to the N7840A
product itself for the latest, most accurate descriptions of the tests.
Source Power Accuracy Test
Function of the Test: To confirm the accuracy of the source output power of your network
analyzer over its full frequency range.
Specification Tested: Test Port Output–Power Level Accuracy
Equipment Used: A power meter, power sensors, and adapters.
Description of the Test:
1. The analyzer's output power level is set to 0 dBm.
2. A power sensor is connected to Port 1.
3. The output power is measured at hundreds of CW frequencies across the analyzer's
frequency range and the values compared to the setting of 0 dBm.
If the Analyzer Fails this Test:
• Perform the
"Source Calibration Adjustment" on page 3-47
• If the analyzer still fails this test, replace the source assembly and repeat this test.
Refer to
"Removing and Replacing the A12 Source 20" on page
Service Guide E8364-90026
"Required Service Test Equipment" on page
Performance Tests (Agilent N7840A Software Package)
and repeat this test.
7-22.