Table 7-1
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
PCA CPI DSP
No adjustment needed
Card
Removable
Calibration of the Touch Screen.
Hard Disk
Drive
No adjustment needed
Power Supply
Assembly
LCD TPANEL
Calibration of the Touch Screen
Assembly
Chapter 7
Post-Repair Procedures
Post-Repair Procedures
Verification
Inspect the Booting Process
Perform "To Execute the diagnostics Test" on
page 187
Perform "To Check the Video Output" on page
200
Inspect the Booting Process
Inspect the Booting Process
Inspect the Booting Process
Perform "To Execute the diagnostics Test" on
page 187
Perform "To Check the LCD" on page 199
331