Performance Test Failure Troubleshooting - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly

Performance test failure troubleshooting

Table 4-7 represents the relationships between the failed test and probable faulty assembly.
If the performance test failure cannot be removed by a proper adjustment, replace the
assembly shown in this table.
Note that this table lists some typical cases. There are possibilities that other assembly may
be faulty. To troubleshoot further, perform the diagnostics test procedures.
NOTE
When Crosstalk, System dynamic range or Uncorrected system performance test fails,
check first whether the connections of the RF semi-rigid cables between the suspicious
assembly and others are tight or loose. Also check for possible disconnection (impairment)
of the cables and connectors.
Table 4-7
Performance test failure troubleshooting information
First failed test
Probable faulty board assembly or parts
OCXO
AUX Input
Frequency Accuracy
(Standard)
Frequency Accuracy (1E5)
#
RF Output Level Accuracy
& Flatness
RF Output Level Linearity
Trace Noise (S11)
Trace Noise (S22)
Trace Noise (S33)
Trace Noise (S44)
Trace Noise (S21, S12)
Trace Noise (S43, S34)
Crosstalk (S21, S12)
Crosstalk (S32)
Crosstalk (S43)
Crosstalk (S14)
System Dynamic Range
(S21)
Chapter 4
BAS
SYN
SYN
VNR
-SRC
-LCL
###
#
###
##
#
###
##
##
###
##
###
##
##
#
##
##
#
##
##
#
##
##
#
##
##
#
##
##
#
#
#
#
#
#
#
#
#
#
#
#
#
##
##
#
Performance test failure troubleshooting
RCV
RCV
RCV
RCV
-1
-2
-3
-4
###
###
###
###
###
###
###
###
##
##
##
##
##
##
##
##
###
Troubleshooting
T2
CBL
-SW
#
#
#
#
#
#
#
#
#
#
###
###
###
###
###
###
###
###
#
203

Advertisement

Table of Contents
loading

Table of Contents