Agilent Technologies E5071C Service Manual page 49

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Name
A1
A2
Power
setting
[dB]
Adjust source power for (Pm [db] @ DA80a) - ( Tx [dB] @ DA80a) + ( Tx [dB] @ DA80)
DA80b
0
80
DA85
5
80
(Tx @ DA85) - ( Tx @ DA80b) - (Pm @ DA85) + (Pm @ DA80b) →
True D.A. = (Running D.A. [dB] @ DA85) + (True D.A. [dB] @ DA80)
DA90
10
80
( Tx @ DA90) - ( Tx @ DA80b) - (Pm @ DA90) + (Pm @ DA80b) True
D.A. = (Running D.A. [dB] @ DA90) + (True D.A. [dB] @ DA80)
DA90a
0
90
Adjust source power for (Pm [db] @ DA90a) - ( Tx [dB] @ DA90a) + ( Tx [dB] @ DA90)
DA90b
0
90
DA95
5
90
( Tx @ DA95) - ( Tx @ DA90b) - (Pm @ DA95) + (Pm @ DA90b)→
True D.A. = (Running D.A. [dB] @ DA95) + (True D.A. [dB] @ DA90)
DA100
10
90
( Tx @ DA100) - ( Tx @ DA90b)- (Pm @ DA100) + (Pm @ DA90b)→
True D.A. = (Running D.A. [dB] @ DA100) + (True D.A. [dB] @ DA90)
DA10
10
0
DA5
5
0
( Tx @ DA5) - ( Tx @ DA10) - (Pm @ DA5) + (Pm @ DA10)
DA0
0
0
( Tx @ DA0) - ( Tx @ DA10) - (Pm @ DA0) + (Pm @ DA10) →
DA+5
0
0
( Tx @ DA+5) - ( Tx @ DA10) - (Pm @ DA+5) + (Pm @ DA10) →
True D.A. = (Running D.A. [dB] @ DA+5) + (True D.A. [dB] @ DA10)
DA+10
0
0
( Tx @ DA+10) - ( Tx @ DA10) - (Pm @ DA+10) + (Pm @ DA10) →
True D.A. = (Running D.A. [dB] @ DA+10) + (True D.A. [dB] @ DA10)
Chapter 2
E5071C Performance Test Calculation Sheet
Pm [dB]
Tx [dB] @ Port (Port Char
OFF)
Performance Test
Running
True
D.A.
D.A.
REF80
REF80
———
———
———
———
———
———
REF90
REF90
———
———
———
———
REF10
REF10
———
———
———
———
———
———
———
———
49