Figure 6-31
LCD Removal
e5071cse1114

Replacement Procedure

Step 1. Replace the LCD by inverse procedure of removal.
NOTE
Fasten the TORX T15 screws (item 5) using a torque screwdriver (set to 1.02 N-m / 9.0
lb-in).
Step 2. You have to calibrate the touch screen. Follow the procedure described on "Calibration of
the Touch Screen" on page 272.
Chapter 6
1
2
3
4
1
Replacement Procedure
LCD Replacement
5
6
319