Control Touch Panel Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver

Removal Procedure

Refer to Figure 6-28 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 266.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 267.
Step 3. Disconnect the cables (item 1, 2) connected through the cover.
Step 4. Remove the two TORX T10 screws (item 3) fastening the control touch panel.
Figure 6-29
Control Touch Panel Replacement
e5071cse1166

Replacement Procedure

Step 1. Replace the control touch panel by inverse procedure of removal.
Chapter 6
Replacement Procedure
Control Touch Panel Replacement