8. Dynamic Accuracy Test

Make a copy of the following calculation sheet for each of the S-parameters listed below:
Option 230, 235, 240, 245, 280 and 285: S21 and S12
Option 430, 435, 440, 445, 480 and 485: S21, S12, S31 and S41
NOTE
A dynamic accuracy calculation example is shown at the end of the calculation sheets.
S-Parameter: _____________ (10 dBm to -100 dBm)
Name
A1
DA10
0
DA15
5
( Tx @ DA15) - ( Tx @ DA10) - (Pm @ DA15) + (Pm @ DA10) →
DA20
10
( Tx @ DA20) - ( Tx @ DA10) - (Pm @ DA20) + (Pm @ DA10) →
DA20a
0
Adjust source power for (Pm [db] @ DA20a) - ( Tx [dB] @ DA20a) + ( Tx [dB] @ DA20)
DA20b
0
DA25
5
( Tx @ DA25) - ( Tx @ DA20b) - (Pm @ DA25) + (Pm @ DA20b) →
True D.A. = (Running D.A. [dB] @ DA25) + (True D.A. [dB] @ DA20)
DA30
10
( Tx @ DA30) - ( Tx @ DA20b) - (Pm @ DA30) + (Pm @ DA20b) →
True D.A. = (Running D.A. [dB] @ DA30) + (True D.A. [dB] @ DA20)
DA30a
0
Adjust source power for (Pm [db] @ DA30a) - ( Tx [dB] @ DA30a) + ( Tx [dB] @ DA30)
DA30b
0
DA35
5
( Tx @ DA35) - ( Tx @ DA30b) - (Pm @ DA35) + (Pm @ DA30b) →
True D.A. = (Running D.A. [dB] @ DA35) + (True D.A. [dB] @ DA30)
DA40
10
( Tx @ DA40) - ( Tx @ DA30b) - (Pm @ DA40) + (Pm @ DA30b) →
True D.A. = (Running D.A. [dB] @ DA40) + (True D.A. [dB] @ DA30)
DA40a
0
Adjust source power for (Pm [db] @ DA40a) - ( Tx [dB] @ DA40a) + ( Tx [dB] @ DA40)
DA40b
0
Chapter 2
A2
Power
Pm [dB]
setting
[dB]
10
10
10
20
20
20
20
30
30
30
30
40
40
E5071C Performance Test Calculation Sheet
Tx [dB] @ Port (Port Char
OFF)
Performance Test
Running
True
D.A.
D.A.
REF
REF
———
———
———
———
———
———
REF20
REF20
———
———
———
———
———
———
REF30
REF30
———
———
———
———
———
———
REF40
REF40
47