Performance Test
E5071C Performance Test Calculation Sheet

4. RF Output Level Linearity Test

NOTE
Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
Option 230, 430:
Option 235, 435:
Option 240, 440:
Option 245, 445:
Option 280, 480:
Option 285, 485:
CW frequency: ______________ Hz
Power level
[dBm]
0
10
7.5
5
2.5
-2.5
-5
-7.5
-10
-12.5
-15
-17.5
-20
NOTE
Make a copy of the following calculation sheet for each combination of the test frequencies
listed below:
Option 280, 285, 480, 485: 6 GHz
44
9 kHz, 100 kHz, 50 MHz, 3 GHz
100 kHz, 50 MHz, 3 GHz
9 kHz, 100 kHz, 50 MHz, 3 GHz, 4.5 GHz
100 kHz, 50 MHz, 3 GHz, 4.5 GHz
9 kHz, 100 kHz, 50 MHz, 3 GHz, 5 GHz
100 kHz, 50 MHz, 3 GHz, 5 GHz
Power meter reading [dBm]
[a]
REF
——————
——————
——————
——————
——————
——————
——————
——————
——————
——————
——————
——————
——————
Test result equation
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF
a-REF