Required Test Equipment; To Execute The Diagnostics Test - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Table 4-1

Required Test Equipment

Table 4-2 shows the equipment required for performing the diagnostic test.
Table 4-2

To Execute the diagnostics Test

To isolate faulty board assembly in analog section, execute the diagnostics test in
accordance with the following procedure. The test procedure needs to be performed using a
mouse in addition to the front panel keys.
NOTE
To perform the diagnostics test properly, the following conditions must be met:
1. Environmental temperature: 23°C ± 5°C
2. Warm-up status indicator in the instrument status bar exhibits "Ready" in blue.
NOTE
Do not operate front panel keys, keyboard and mouse during the diagnostics test. Changing
the instrument settings while the diagnostics test is in progress will cause incorrect test
results.
Step 1. Connect a mouse to the rear panel USB connector
Step 2. Press
Step 3. Press
shown in Figure 4-9.
Chapter 4
Diagnostic test group menu
Number
Test Name
2
INTERNAL LEVEL MONITOR TEST
3
RF OUTPUT LEVEL POWER SEEP
4
RF OUTPUT LEVEL RANGE TEST
5
RECEIVER ABSOLUTE MEASUREMENT TEST
6
RECEIVER COMPRESSION TEST
7
RECEIVER IF RANGING TEST
Required Equipment
Required Test Equipment
24 inch 50 Ohm cable
key.
to select
Load Project
Load Project
Troubleshooting Using Diagnostics Test
Qty
Recommended Model
1
8120-8862
function."Open" dialog box will be displayed as
Troubleshooting
187