Table 7-1
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
Synthesizer
Perform the following required
Module
adjustment.
- Synthesizer Gain
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
Receiver
Perform the following required
Module
adjustments.
- Source Output Power
- Receiver IF Range
- Receiver Port Characteristics
- Receiver Absolute Gain
Level Vernier
Perform the following required
Module
adjustments.
- Source Output Power
- Receiver Absolute Gain
Chapter 7
Post-Repair Procedures
Post-Repair Procedures
Verification
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Dynamic accuracy test
- Uncorrected system performance
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Dynamic Accuracy test
- Uncorrected System Performance Test