Agilent Technologies E5071C Service Manual page 273

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Figure 6-5
Touch Panel Calibration Screen
e5071cse1078
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE
With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Chapter 6
Replacement Procedure
Removable Hard Disk Drive Replacement
273