Replacement Procedure

Digital Bridge Board Replacement

Digital Bridge Board Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to Figure 6-12 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 266.
Step 2. Remove the three TORX T10 screws (item 1) fastening the angle.
Step 3. Disconnect the flat cables (item 2) from the digital bridge board.
Step 4. Remove the two TORX T10 screws (item 3) fastening the board to the rear panel.
Step 5. Lift the digital bridge board (item 4).
Step 6. Disconnect the cables (item 5) from the digital bridge board.
Step 7. Remove the two TORX T10 screw (item 6) fastening the angle.