Post-Repair Procedures
Post-Repair Procedures
Table 7-1
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
Distributer
Perform the following required
Module
adjustments.
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
T2 Switch
Perform the following required
adjustments
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
CPU Mother
Perform "To configure the CPU Mother
board
Board Assembly and BIOS" on
Assembly
page 208.
PCA Front
No adjustment needed
Key
Handler I/O
No adjustment needed
Board
GPIB Board
No adjustment needed
No adjustment needed
USB Interface
Card
330
Verification
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Dynamic Accuracy test
- Uncorrected System Performance Test
Inspect the Booting Process
Perform "To Execute the diagnostics Test" on
page 187
Perform "To Check the Front Panel" on page 198
Perform "To Execute the diagnostics Test" on
page 187
Perform "To Check the GPIB" on page 200
Perform "To Check the USB Interface Card" on
page 200