Post-Repair Procedures - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Post-Repair Procedures

Post-Repair Procedures
Post-Repair Procedures
Table 7-1 Post Repair Procedures lists the required procedures that must be performed
after the replacement of an assembly or CPU. These are the recommended minimum
procedures to ensure that the replacement is successfully completed.
It is recommended to confirm the performance executing with all of Adjustment and
performance test as final step.
Table 7-1
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
Analog I/F
Perform the following required
Board
adjustment.
- Write ID
- Save to Flash memory
Reference
Perform the following required
Oven Board
adjustment.
- OCXO
Analog Base
Perform the following required
Module
adjustment.
- AUX Input
- Frequency Reference
328
Verification
Perform "To Execute the diagnostics Test" on
page 187
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Dynamic Accuracy Test
- Uncorrected System Performance Test
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- Frequency Accuracy Test
Perform "To Execute the diagnostics Test" on
page 187
Perform the follwoing performance test.
- AUX Input Test
- Frequency Accuracy Test
Chapter 7