Replacement Procedure

LCD Replacement

LCD Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver

Removal Procedure

Refer to Figure 6-31 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 266.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 267.
Step 3. Remove the front panel cover as described in "Front Panel Cover Removal" on page 314.
Step 4. Remove the inverter board as described in "Inverter Board Replacement" on page 316.
Step 5. Release the clamped cables from the clamps (item 1) on the LCD.
Step 6. Disconnect the cable (item 2) from the probe power.
Step 7. Remove the two TORX T10 screws (item 3) fastening the probe power and lift it (item 4).
Step 8. Remove the two TORX T15 screws (item 5) fastening the LCD to the front panel.
Step 9. Disconnect the cable (item 6) from the front panel keyboard.