Performance Test
E5071C Performance Test Record
Without Option 1E5
Frequency [Hz]
Test limit [Hz]
100 k
50 M
134.1 M
548 M
3 G
4.5 G
With Option 1E5
Frequency [Hz]
Test limit [Hz]
100 k
50 M
134.1 M
548 M
3 G
4.5 G
Option 280, 480
Without Option 1E5
Frequency [Hz]
Test limit [Hz]
9 k
100 k
50 M
134.1 M
548 M
3 G
8.5 G
54
Test result [Hz]
± 0.5
± 250
± 670.5
± 2.74 k
± 15 k
± 22.5 k
Test result [Hz]
± 0.1
± 50
± 134.1
± 0.548 k
± 3 k
± 4.5 k
Test result [Hz]
± 0.045
± 0.5
± 250
± 670.5
± 2.74 k
± 15 k
± 42.5 k
Measurement
uncertainty [Hz]
± 0.004
± 1.7
± 4.6
± 16.4
± 90
± 582
Measurement
uncertainty [Hz]
± 0.003
± 0.58
± 1.6
± 6.4
± 35
± 52.5
Measurement
uncertainty [Hz]
± 0.0004
± 0.004
± 1.7
± 4.6
± 16.4
± 90
± 254