Receiver Module Replacement For 4-Port Model; Tools Required; Removal Procedure - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Replacement Procedure

Receiver Module Replacement for 4-Port model

Receiver Module Replacement for 4-Port model

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)

Removal Procedure

Refer to Figure 6-18 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 266.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 267.
Step 3. Disconnect the two semi-rigid cable (item 1).
Step 4. Disconnect the two semi-rigid cable (item 2).
Step 5. Disconnect the four semi-rigid cables (item 3).
Step 6. Remove the four TORX T10 screws fastening the module to the bottom. (item 4)
Step 7. Slide the receiver module (item 5).
302
Chapter 6

Advertisement

Table of Contents
loading

Table of Contents