Replacement Procedure

Handler I/O Board Replacement

Handler I/O Board Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to Figure 6-13 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 266.
Step 2. Disconnect the flat cable (item 1) from the handler I/O board.
Step 3. Remove the two TORX T10 screws (item 2) fastening the handler I/O board.
Step 4. Slide the handler I/O board toward backward (item 3).