Performance Test
E5071C Performance Test Record
Power level [dBm]
Test limit [dB]
-45
-50
-55
-60
-65
-70
-75
-80
-85
-90
-95
-100

Uncorrected System Performance Test

@ Correction: Off, System Correction: On
Option 230
Port 1
System
Frequency
performance
range [Hz]
Directivity
9 k - 300 k
Directivity
300 k -3 G
Source match
9 k - 300 k
Source match
300 k -3 G
Load match
9 k - 300 k
Load match
300 k - 3 G
Transmission
9 k - 300 k
Tracking
Transmission
300 k - 3 G
Tracking
Reflection
9 k - 300 k
Tracking
140
Test result [dB]
± 0.078
± 0.091
± 0.106
± 0.125
± 0.151
± 0.189
± 0.248
± 0.346
± 0.509
± 0.785
± 1.248
±2.008
Test limit
Test result
[dB]
[dB]
< -20
< -25
< -20
< -25
< -12
< -17
± 1.5
± 1.0
± 1.5
Measurement
uncertainty [dB]
± 0.021
± 0.021
± 0.024
± 0.024
± 0.026
± 0.025
± 0.011
± 0.014
± 0.021
± 0.033
± 0.069
± 0.060
Measurement
uncertainty [dB]
0.09
0.81
0.05
3.00
0.04
0.59
± 0.03
± 0.20
± 0.04