Dynamic Accuracy Test - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Dynamic Accuracy Test

@ Reference power level: -10 dBm
Direction: S21
Power level [dBm]
Test limit [dB]
10
5
0
-5
-15
-20
-25
-30
-35
-40
-45
-50
-55
-60
-65
-70
-75
-80
-85
-90
-95
-100
Direction: S12
Power level [dBm]
Test limit [dB]
10
5
Chapter 2
E5071C Performance Test Record
Test result [dB]
± 0.207
± 0.075
± 0.041
± 0.031
± 0.030
± 0.035
± 0.040
± 0.045
± 0.056
± 0.067
± 0.078
± 0.091
± 0.106
± 0.125
± 0.151
± 0.189
± 0.248
± 0.346
± 0.509
± 0.785
± 1.248
±2.008
Test result [dB]
± 0.207
± 0.075
Performance Test
Measurement
uncertainty [dB]
± 0.010
± 0.010
± 0.007
± 0.009
± 0.006
± 0.007
± 0.010
± 0.011
± 0.015
± 0.019
± 0.021
± 0.021
± 0.024
± 0.024
± 0.026
± 0.025
± 0.011
± 0.014
± 0.021
± 0.033
± 0.069
± 0.060
Measurement
uncertainty [dB]
± 0.010
± 0.010