7. DYNAMIC ACCURACY TEST

Description

This test checks the dynamic accuracy of the E5071C. The dynamic accuracy is tested at
1.195 GHz for each receiver port using the Agilent Z5623A (with option H01) Dynamic
Accuracy Test Kit. The 1.195 GHz source signal of the E5071C goes through the Z5623A's
first attenuator variable from 0 dB to 11 dB in 1 dB steps and second attenuator variable up
to 100 dB in 10 dB steps and is measured with the receiver port. The test procedure checks
the measured receiver-input power for each 5 dB increment in the range from 10 dBm to
100 dBm. The ratios of the measured receiver-input powers to the reference input level of
-10 dBm are calculated and compared to the attenuation values of the Z5623A. The 1.195
GHz signal flow is split with a power splitter present between the two attenuators in the
Z5623A and measured with an external power meter to calibrate the attenuated power
level.
NOTE
Since the dynamic accuracy does not have frequency dependence, this test is performed at
1.195 GHz only.

Test Equipment

Power meter
Power sensor
Dynamic Accuracy Test Kit
6 dB fixed attenuator
Coaxial cable with N-type (m) connectors, 61 cm
(24 in), 2 ea.
GPIB controller (or PC with GPIB Interface Card)
GPIB Interconnection Cable
NOTE
Before starting this test, input and store the calibration data of 8482A to E4419A/B.
Chapter 2
Performance Test
7. DYNAMIC ACCURACY TEST
Agilent E4419A/B
Agilent 8482A
Agilent Z5623A w/Opt. H01
Agilent 8491A w/Opt. 006 and
H60
Agilent N6314A (p/n 8120-8862)
Agilent 10833A/B
37