NOTE
It is recommended that a copy of the test record on page 26 be made, and the values be
recorded on the copy, thus preserving the original for future use.
Test Procedure
NOTE
This procedure assumes the use of the recommended equipment model numbers. The
actual steps required, therefore, may differ for other model numbers of equipment used.
Step 1. Setup the equipment as shown in Figure 1-2.
a. Connect the DC REFERENCE OUTPUT connector on the rear panel of the range
calibrator to the DVM voltage input.
b. Connect the POWER METER output of the range calibrator to the input of the power
meter being tested.
c. Switch on the power to the power meter, the range calibrator, and the digital voltmeter.
Figure 1-2
Setup for the Power Meter Accuracy Test
Step 2. Press the power meter: Press
Step 3. Perform the following steps for each channel on the power meter:
a. Set to read in dBm: Press
b. Set the ref cal factor to 100%: Press
necessary.
c. Set the cal factor to 100%: Press
if necessary.
d. Set readout to 0.001 dBm: Press
e. Set filter step detect on and filter length to 512: Press
Input Setting
then set the filter length to
Step 4. Setup the digital voltmeter (DVM) as follows:
a. Reset the DVM: Press the
b. Set the sample period to a value greater than one second: Press
Step 5. Set the range calibrator controls as follows:
C h a p te r 1
, then
[Preset/Local]
, then
[dBm/W]
dBm
[Zero/Cal]
[Frequency/Cal Fac]
[Meas Setup]
,
,
,
[More]
Ch A/B Filter
Step Det On
.
512
followed by
blue key
General Information
Power Meter Accuracy Test
.
Confirm
.
,
,
, then set to
Cal
A/B Ref CF
,
, then set to
A/B Cal Fac
, then
, to highlight
Resolution 1234
, channel
[System/Inputs]
,
,
,
Filter On
Mode MAN
.
Reset
,
,
, then
NPLC
5
0
, if
100.0
,
100.0
.
4
or
A
B
,
Length
.
Enter
23