Agilent Technologies E5071C Service Manual page 206

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Troubleshooting
Performance test failure troubleshooting
Table 4-9
Adjustment failure troubleshooting information
Failed
Failed test item
Adjustment
Synthesizer
vSrcDblDet / vSrcAlcDet
Gain
vLrcDblDet / vLrcAlcDet
Source Output
ccSrcVnr{1|2}Vref
Power
RchGain
ccSrcATT{1|2|
matPowSrc[1000]
matPowSrc[4000]
Receiver IF
ccIFRange[{R1|T1}]
Range
ccIFRange[{21|T2}]
ccIFRange[{R3|T3}]
ccIFRange[{R4|T4}]
Receiver Port
S1,D1,R1
Characteristics
S2,D2,R2
S3,D3,R3
S4,D4,R4
L21,L12,T21,T12
L31,L13,T31,T13
L41,L14,T41,T14
L32,L23,T32,T23
L42,L24,T42,T24
L43,L34,T43,T34
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
206
Probable faulty board assembly or parts
###
###
###
#
#
#
##
###
###
###
#
#
#
#
###
#
#
#
###
#
###
#
###
#
###
#
##
##
##
##
##
##
##
##
##
##
##
##
##
##
##
##
#
##
##
##
##
##
##
###
###
###
###
#
###
#
###
#
###
#
###
#
###
#
###
Chapter 4

Advertisement

Table of Contents
loading