5. If you want to modify the temperature steps, refer to these steps:
To edit or delete a temperature step, tap the More
To group the temperature steps, check the corresponding boxes of more than one
temperature step, then tap Group.
To ungroup a group of temperature steps, tap the More
6. Perform the following steps in the Imaging tab:
6a.
In the Imaging tab, select the applicable channel, then enter the exposure duration and
gain in the Exposure duration and Gain fields.
6b.
To include more steps in the run, tap the Add
step. Provide the required information for the step. Repeat this step if more steps are
needed for the run. In total, 9 steps can be performed per plate.
6c.
Tap Save to save your progress or tap Done to save the run and go back to the Running
Status window.
Note: If any required field is not completed, an error message displays to list the missing
information required in each field.
QIAcuity User Manual 06/2022
icon, then tap Edit or Delete.
icon, then tap Ungroup.
icon, then select the applicable
80